MENB1040
Universal 40 Watt Series
Medical Switch-Mode Power Supply
•100-240VAC Universal Input
•Meets EISA2007, CEC Efficiency Level V,
EU (EC) No 278/2009 Phase II
•Desktop Style
•Certified to UL/EN60601-1
•5V to 24V Single Output Models, up to 40W
•Modified and Custom Designs Available
•Regulated Output with Low Ripple
•Impact-Resistant Polycarbonate Enclosure
•No load Power Consumption <0.3W
RoHS
3 Year Warranty
Specifications
All Specifications are typical at nominal input, full load at 25°C unless otherwise stated.
AC Input
100-240VAC, -/-10%, 47-63 Hz, 1∅ MTBF
>100,000 hours (calculated)
Input Current
100VAC: 1.1A Hold-up Time
18 ms min. @ 115Vac, 60 ms @ 230 Vac
Inrush Current
60 A peak, 264 V, cold start Overload Protection
Hiccup Mode
Input Fuse
3.15A, 250V Internal Primary Current Fuse is provided Short Circuit Protection
Hiccup Mode
Efficiency
Output Voltage
Meets EISA2007, CEC Efficiency Level V,
EU (EC) No 278/2009 Phase II
See chart
Topology
Safety Standards
Switching – Fixed Frequency Flyback
EN/IEC/CSA/UL60601-1
Output Power
See chart EMC
See chart below
Ripple and Noise
1% pk-pk max., 20MHz BW Dielectric Withstand
Line & Load Voltage Regulation
Line: +/- 1%, Load: +/-5%
Transient Response
500µs max., 50% load step, typical
Minimum Load
Not required
Case Material
Black 94V0 Polycarbonate
Case Dimensions
102 x 60 x 32mm. See outline drawing
Weight
250g
Operating Temperature
Storage Temperature
Relative Humidity
Altitude
Output Cable
Output Connector
Input-Output: 5,656Vdc
Input-GND: 1,500 Vac, Output-GND: 500Vdc
0° to 40°C, no derating
-30 to +85°C
5% to 95%, non-condensing
0 to 10,000 ft
#18AWG (UL1185), 1,500mm, 2 conductor
2.5mm barrel type (Ault #3), center positive (+)
Conducted Emissions
Radiated Emissions
Line Frequency Harmonics
Voltage Fluctuations/Flicker
Static Discharge Immunity
Radiated RF Immunity
EFT/Burst Immunity
Line Surge Immunity
Conducted RF Immunity
Power Frequency Magnetic Field Immunity
Voltage Dip Immunity
EMC Specifications
EN55011 Class B, FCC Part 15, Class B.
EN55011 Class B, FCC Part 15, Class B.
EN61000-3-2, Class A
EN61000-3-3
EN61000-4-2, 6kV Contact Discharge, 8kV air discharge
EN61000-4-3, 3V/m.
EN61000-4-4, 2kV/5kHz.
EN61000-4-5, 1kV differential, 2kV common-mode
EN61000-4-6, 3Vrms
EN61000-4-8, 3A/m
EN61000-4-11, Criteria A