GPFM600 Datasheet, PDF (1/2 Pages) SL Power Electronics – 600 Watt Global Performance Switchers
GPFM600 Medical
600 Watt Global Performance Switchers
FEATURES:
• Small package (12.00” x 5.50” x 3.09”)
• Power factor corrected to EN61000-3-2 Class A
• Full load rating to 60°C
• 85 to 264 Vac Input
• Medically Approved to UL2601-1, IEC601-1, CSA
C22.2 No. 601.1.2
• 2 year warranty
•
marked to LVD
SPECIFICATIONS:
Ac Input
85-264 Vac, 47-63 Hz single phase.
Output Power
Continuous output power, 600 watts.
Hold-upTime
20 ms minimum from loss of ac input until output reaches
3% below nominal.
Line and Load Regulation
±1% over the entire input voltage and frequency range,
and over the entire output current range.
Minimum Load
No minimum load required.
Overload Protection
Current limit 105% minimum, 125% maximum of the
specified full load current.
Output Noise
50 mV pk-pk maximum.
Transient Response
Maximum of 5% change in Vout for a load change of 1.25 A.
Recovers to ±1% within 500u sec.
Overvoltage Protection
Built in on all models. OVP crowbar reduces output
voltage below nominal rating in less than 50 ms. Output
voltage decay is dependent upon loading.
DC Power Good
Optocoupler isolated output indicates that the DC output
is equal to or greater than 21.6 (-0 + 0.25) volts. This
signal shall be driven when in the active ‘DC Good’ state.
Efficiency
Minimum of 75% under full load and input of 120 Vac 60 Hz.
Inrush Current
Limit inrush current to 125% of maximum operating
current. Turn on time of the power supply not to exceed 3
seconds over the full input voltage and frequency range.
EMI/EMC Compliance
All models include built-in EMI filtering to meet emissions
requirements:
EMI SPECIFICATIONS
COMPLIANCE LEVEL
Conducted Emissions
EN55011 Class B; FCC Class B
Static Discharge
EN61000-4-2, 6 kV contact, 8 kV air
RF Field Susceptibility
EN61000-4-3, 3 V/meter
Fast Transients/Bursts
EN61000-4-4, 2 kV, 5 kHz
Surge Susceptibility
EN61000-4-5, 1 kV diff., 2 kV com.
Vibration and Shock
All models are to meet MIL-STD 810E, method 514.4
Category.
Environmental
Designed for 0 to 60°C operation at full rated output
power; derate output current and total output power by
2.5% per °C above 50°C.
Medical Leakage Current
150 µA under normal conditions. Maximum under single
fault conditions, 250 µA. Patient sink leakage with 120
Vac, 60 Hz applied to any output with respect to the
chassis will be less than 500 µA.
Medical Safety Agency
Approved to UL2601-1, EN60601-1-2, CSA C22.2 No.
601.1.2.