G2T60 MEDICAL SERIES SINGLE OUTPUT  POWER SUPPLY
Low Power Medical Power Supplies 60 Watt AC/DC Universal Input Power Supply
• 100 - 240 Vac nominal input range
• Overvoltage and short circuit protection
• Approved to UL, CSA and EN Standards
• Class I input available (GNT60)
SUMMARY
The G2T60 Series is a 60 W universal input AC/DC power
supply in a very small footprint. With the medical approvals
the G2T60 is ideal for a variety of medical device applications
including, small single board computers, battery charging, and
running small motors, pumps, and solenoids. The series with
full approval to EN60601-1 and EN60950-1 Standard, improves
design-in time and reduces end system compliance costs.
All specifications are typical at nominal input, full load at 25 OC unless otherwise stated
SPECIFICATIONS
OUPUT SPECIFICATIONS
Output Power
Natural Convection
Total Regulation
Rise time
At turn-on
Transient response
Main output 50%
0.2 A/ms
Temperature Coef-
ficient
Overvoltage protection
Short Circuit protection Power cycling
60 Watts max.(65 W with airflow)
See table
2.0 s max.
5% max. dev. 1 ms recovery to
1%
+/-0.03%/oC
125% +/-10%
Yes
INPUT SPECIFICATIONS
Input Voltage range
Universal input
Input frequency
Input surge current
Ground Leakage current
Thermistor limited
132 Vac 60 Hz
Input current
Input fuse
120 Vac
230 Vac
F1,F2
100-240 nom (85 - 264 max)
Vac
55 Hz +/- 10 Hz
37 A max.
100 mA Input to grounded-
output
1.2 A
0.7 A
3.15 A
EMC INFORMATION
RF emissions
EN55011
Line freq. harmonics
Voltage fluctuations
ESD
EN61000-3-2
EN61000-3-3
EN61000-4-2
Radiated immunity
EN61000-4-3
Fast Transients (EFT) EN61000-4-4
Conducted Class B
Radiated Class A
Complies
Complies
Contact 6 kV
Air 8 kV
3 V/m
2 kV
Line surge immunity
EN61000-4-5 DIfferential 1 kV
Common 2 kV
Conducted immunity EN61000-4-6 3 Vrms
Power freq. mag. field EN61000-4-8 3 A/m
Voltage dip immunity EN61000-4-11 95% dip, 0.5 cycle 60% dip, 5 cycles
30% dip, 25 cycles
Note: Voltage dip immunity at 100 V is done with a 20% load
ENVIRONMENTAL SPECIFICATIONS
Thermal Performance
operating ambient
(see ratings chart)
non-operating
0 - 50 oC
Convection cooled
50 -70 oC ambient
Convection cooled
Relative Humidity
non-condensing
Maximum Altitude
operating / non-operating
Vibration
Shock
5 Hz- 500 Hz
per MIL-STD-810E
0-70 oC
-40 to +85 oC
60 Watts
derate to 50%
5% - 95% RH
10,000 ft. / 40,000
ft. max.
2.5 g rms
516.4 part IV