English
Language : 

SCX01DN Datasheet, PDF (2/10 Pages) List of Unclassifed Manufacturers – Precision Compensated Pressure Sensors
SCX Series
Precision compensated pressure sensors
PRESSURE SENSOR CHARACTERISTICS1
STANDARD PRESSURE RANGES
Part Number
SCX01DN
SCX05DN
SCX15AN
SCX15DN
SCX30AN
SCX30DN
SCX100AN
SCX100DN
SCX150DN
Operating pressure
0 - 1 psid
0 - 5 psid
0 - 15 psia
0 - 15 psid
0 - 30 psia
0 - 30 psid
0 - 100 psia
0 - 100 psid
0 - 150 psid
Proof pressure2
20 psid
20 psid
30 psia
30 psid
60 psia
60 psid
150 psia
150 psid
150 psid
Full-scale span1,3
Min.
17.82 mV
59.4 mV
89.1 mV
89.1 mV
89.1 mV
89.1 mV
99.0 mV
99.0 mV
89.0 mV
Typ.
18 mV
60 mV
90 mV
90 mV
90 mV
90 mV
100 mV
100 mV
90 mV
Max.
18.18 mV
60.6 mV
90.9 mV
90.9 mV
90.9 mV
90.9 mV
101.0 mV
101.0 mV
91.0 mV
Maximum ratings (for all devices)
Supply voltage VS
Common-mode pressure
Lead temperature (soldering, 4 seconds)
+20 VDC
50 psig
250°C
Environmental specifications (for all devices)
Temperature range
Compensated
Operating
Storage
Humidity limits (no condensation)
0 to 70°C
-40°C to +85°C
-55°C to +125°C
0 to 99 %RH
COMMON PERFORMANCE CHARACTERISTICS1
Characteristic
Zero pressure offset12
Combined linearity and hysteresis4
Temperature effect on span (0 - 70°C)5
Temperature effect on offset (0 - 70°C)5
Repeatability6
Input impedance7
Output impedance8
Common-mode voltage9
Response time10
Long term stability of offset and span11
Min.
-300
---
---
---
---
---
---
5.8
---
---
Typ.
0
±0.1
±0.2
±100
±0.2
4.0
4.0
6.0
100
±0.1
Max.
+300
±0.5
±1.0
±500
±0.5
---
---
6.2
---
---
Unit
µV
%FSO
%FSO
µV
%FSO
kΩ
kΩ
VDC
µsec
mV
Specification notes:
1.
Reference conditions: unless otherwise noted: supply voltage, V = 12 V, T = 25°C, common-mode line pressure = 0 psig,
S
A
pressure applied to Port B. For absolute devices only, pressure is applied to Port A and the output polarity is reversed.
2. Maximum pressure above which causes permanent sensor failure.
3. Span is the algebraic difference between the output voltage at full-scale pressure and the output at zero pressure. Span
is ratiometric to the supply voltage.
4. See Definition of Terms. Hysteresis - the maximum output difference at any point within the operating pressure range for
increasing and decreasing pressure.
5. Maximum error band of the offset voltage and the error band of the span, relative to the 25°C reading.
6. Maximum difference in output at any pressure with the operating pressure range and temperature within 0°C to +50°C after:
a) 1,000 temperature cycles, 0°C to +70°C
b) 1.5 million pressure cycles, 0 psi to full-scale span
7. Input impedance is the impedance between pins 2 and 4.
8. Output impedance is the impedance between pins 3 and 5.
9.
This is the common-mode voltage of the output arms (pins 3 and 5) for V = 12 V .
S
DC
10. Response time for a 0 psi to full-scale span pressure step change, 10 % to 90 % rise time.
11. Long term stability over a one year period.
12. Maximum zero pressure offset for absolute devices is 0 ±500 µV.
2/10
August 2005 / 053
www.sensortechnics.com