English
Language : 

501907 Datasheet, PDF (1/2 Pages) Sensortechnics GmbH – 10mm² PIN detector
First Sensor Ionizing Radiation Detector Data Sheet
Part Description X10-6 TO
Order # 501907
Version 16-12-10
Features
• 10mm² PIN detector
• Low dark current
• Low capacitance
• Light blocking glob top
Description
PIN detector for ionizing radiation
with 10mm² active area. TO package
with epxoy based light blocking glob
top. Non hermetic.
Application
• X-Ray detector
• Gamma radiation detector
RoHS
2002/95/EC
Absolute maximum ratings
Symbol
Parameter
Min Max Unit
TSTG Storage temp
TOP Operating temp
UR Reverse voltage
-20 80 °C
0 40 °C
10
V
IPEAK Peak DC current
10 mA
Schematic
PIN 3
Dark current as fct of reverse bias (23 °C)
1,00E-09
1,00E-10
PIN 1
1,00E-11
0
5
10
Reverse bias (V)
Electro-optical characteristics @ 23 °C
Symbol
Characteristic
Test Condition
No of elements
Active area
Window material
ID Dark current
C Capacitance
U = 10 V
U=0V
C Capacitance
U = 10 V
tR Rise time
UBR Breakdown voltage
Energy range of
U = 10 V; 20 keV; RL = 50 Ω
IR = 2 µA
X-ray radiation
detectable radiation
Gamma radiation
Min
Typ
Max Unit
1
10 mm², 3,57 mm diameter
Glob top: black epoxy
0.5
nA
100
pF
18
pF
20
ns
30
V
5
keV
5
keV
European, International Sales:
First Sensor AG
Peter-Behrens-Strasse 15
12459 Berlin
Germany
T +49 30 6399 2399
F +49 30 639923-752
sales.opto@first-sensor.com
USA:
First Sensor Inc.
5700 Corsa Avenue #105
Westlake Village
CA 91362 USA
T +1 818 706 3400
F +1 818 889 7053
sales.us@first-sensor.com