First Sensor Ionizing Radiation Detector Data Sheet
Part Description X10-É£ TO
Order # 50190401
Version 14-09-12
Foto
Features
• 10 mm² PIN detector
• Very low dark current
• Very low dark capacitance
• Fully depletable
• Guard ring diode
Description
Circular active area PIN photodiode
with 10 mm² active area. Metal can
type non-hermetic isolated TO8Si
package with aluminzed Kapton
window.
Application
• Dosimeter
• Radiation monitoring
• Gamma diffraction
• X-ray fluorescence
• Density measurements
RoHS
2002/95/EC
Absolute maximum ratings
Symbol
Parameter
Min Max Unit
TSTG Storage temp
TOP Operating temp
UR Reverse voltage
-55 125 °C
-20 70 °C
300 V
IPEAK Peak DC current
10 mA
Schematic
Gamma - Absorption (23 °C)
100%
90%
80%
70%
60%
50%
40%
30%
20%
10%
0%
0
5 10 15 20 25 30 35 40 45 50
Gamma energy (keV)
Electro-optical characteristics @ 23 °C
Symbol
Characteristic
Test Condition
No of elements
Active area
Diameter
Window material
Scintillator
ID Dark current
U=0V
U = 120 V (full depletion)
C Capacitance
U=0V
U = 120 V (full depletion)
tR Rise time
UBR Breakdown voltage
Energy range of
U = 120 V; 20 keV; RL = 50 Ω
IR = 2 µA
Alpha radiation
detectable radiation
Beta radiation
X-Ray / Gamma radiation
Min
Typ
Max Unit
1
3.75
mm
double-aluminized Kapton (8 µm)
none
1
2
pA
20
40
pA
65
pF
4.5
pF
15
ns
300
500
V
2
MeV
20
1000 keV
1
50 keV
European, International Sales:
First Sensor AG
Peter-Behrens-Strasse 15
12459 Berlin
Germany
T +49 30 6399 2399
F +49 30 639923-752
sales.opto@first-sensor.com
USA:
First Sensor Inc.
5700 Corsa Avenue #105
Westlake Village
CA 91362 USA
T +1 818 706 3400
F +1 818 889 7053
sales.us@first-sensor.com