English
Language : 

DPO3EMBD Datasheet, PDF (14/22 Pages) List of Unclassifed Manufacturers – Mixed Signal Oscilloscopes
Datasheet
Power measurements (optional)
Safe operating area
Graphical display and mask testing of switching device safe operating area measurements.
dV/dt and dI/dt measurements
Cursor measurements of slew rate.
Waveform math
Arithmetic
Math functions
FFT
Advanced math
Add, subtract, multiply, and divide waveforms.
Integrate, Differentiate, FFT.
Spectral magnitude. Set FFT Vertical Scale to Linear RMS or dBV RMS, and FFT Window to Rectangular, Hamming, Hanning, or
Blackman-Harris.
Define extensive algebraic expressions including waveforms, reference waveforms, math functions (FFT, Intg, Diff, Log, Exp, Sqrt,
Sine, Cosine, Tangent), scalars, up to two user-adjustable variables and results of parametric measurements (Period, Freq, Delay,
Rise, Fall, PosWidth, NegWidth, BurstWidth, Phase, PosDutyCycle, NegDutyCycle, PosOverShoot, NegOverShoot, PeakPeak,
Amplitude, RMS, CycleRMS, High, Low, Max, Min, Mean, CycleMean, Area, CycleArea, and trend plots), e.g.,(Intg(Ch1 -
Mean(Ch1)) × 1.414 × VAR1).
Software
NI LabVIEW SignalExpress™
Tektronix Edition
OpenChoice® Desktop
IVI driver
e*Scope® Web-based remote
control
A fully interactive measurement software environment optimized for your Tektronix oscilloscope, enables you to instantly acquire,
generate, analyze, compare, import, and save measurement data and signals using an intuitive drag-and-drop user interface that
does not require any programming.
Standard support for acquiring, controlling, viewing, and exporting your live analog-channel signal data is permanently available
through the software. The full version (SIGEXPTE) adds additional signal processing, advanced analysis, mixed signal, sweeping,
limit testing, and user-defined step capabilities and is available for a 30-day trial period standard with each instrument.
Enables fast and easy communication between a Windows PC and your oscilloscope using USB or LAN. Transfer and save
settings, waveforms, measurements, and screen images. Included Word and Excel toolbars automate the transfer of acquisition
data and screen images from the oscilloscope into Word and Excel for quick reporting or further analysis.
Provides a standard instrument programming interface for common applications such as LabVIEW, LabWindows/CVI,
Microsoft .NET, and MATLAB.
Enables control of the oscilloscope over a network connection through a standard web browser. Simply enter the IP address or
network name of the oscilloscope and a web page will be served to the browser.
Display system
Display type
9 inch (228.6 mm) wide format liquid crystal TFT color display.
Display resolution
800 horizontal × 480 vertical pixels (WVGA).
Waveform styles
Vectors, Dots, Variable Persistence, Infinite Persistence.
Graticules
Full, Grid, Cross Hair, Frame, IRE and mV.
Format
YT and XY.
Maximum waveform capture rate >50,000 wfm/s.
14 www.tektronix.com