English
Language : 

08S121-S20S3 Datasheet, PDF (2/3 Pages) Rosenberger Hochfrequenztechnik GmbH & Co. KG – Calibration Adaptor
Technical Data Sheet
RPC-1.85
Calibration Adaptor
Plug/Plug
08S121-S20S3
Electrical data
Frequency range
Return loss
DC to 70 GHz
≥ 28 dB, DC to 4 GHz
≥ 20 dB, 4 GHz to 40 GHz
≥ 17 dB, 40 GHz to 70 GHz
Mechanical data
Mating cycles
Maximum torque
Recommended torque
Gauge
≥ 500
1.65 Nm
0.90 Nm
0.00 mm to 0.03 mm
General standard definitions
For proper operation the vector network analyzer (VNA) needs a model describing the electrical behaviour
of this calibration standard. The different models, units, and terms used will depend on the VNA type and they
will have to be entered into the VNA. All values are based on typical geometry and plating.
Offset Zo / Impedance / Zo
Offset Delay
Length (electrical) / Offset Length
Offset Loss
Loss
50 Ω
61.042 ps
18.3 mm
4.00 GΩ/s
0.0212 dB/ GHz
Environmental data
Operating temperature range1
Rated temperature range of use2
Storage temperature range
+20 °C to +26 °C
0 °C to +50 °C
-40 °C to +85 °C
RoHS
compliant
1 Temperature range over which these specifications are valid.
2 This range is underneath and above the operating temperature range, within the calibration adaptor is fully functional
and could be used without damage.
Rosenberger Hochfrequenztechnik GmbH & Co. KG
P.O.Box 1260 D-84526 Tittmoning Germany
www.rosenberger.de
Tel. : +49 8684 18-0
Fax : +49 8684 18-499
Email : info@rosenberger.de
Page