03S150-C10S3 Datasheet, PDF (2/3 Pages) Rosenberger Hochfrequenztechnik GmbH & Co. KG – RPC-3.50 Calibration Load Plug | |||
|
◁ |
Technical Data Sheet
RPC-3.50
Calibration Load
Plug
Electrical data
Frequency range
Return loss
DC Resistance
Power handling
03S150-C10S3
DC to 26.5 GHz
ï³ 40 dB, DC to 4 GHz
ï³ 35 dB, 4 GHz to 8 GHz
ï³ 30 dB, 8 GHz to 26.5 GHz
50 ï ï± 0.25 ï
⤠0.5 W
Mechanical data
Mating cycles
Maximum torque
Recommended torque
Gauge
ï³ 500
1.70 Nm
0.90 Nm
0.00 mm to 0.04 mm
General standard definitions
For proper operation the vector network analyzer (VNA) needs a model describing the electrical behaviour
of this calibration standard. The different models, units, and terms used will depend on the VNA type and they
will have to be entered into the VNA. All values are based on typical geometry and plating.
Offset Zo / Impedance / Zo
Offset Delay
Length (electrical) / Offset Length
Offset Loss
Loss
50 ï
0.0000 ps
0.00 mm
0.00 Gâ¦/s
0.0000 dB/ GHz
Environmental data
Operating temperature range1
Rated temperature range of use2
Storage temperature range
+20 °C to +26 °C
0 °C to +50 °C
-40 °C to +85 °C
RoHS
compliant
1 Temperature range over which these specification are valid.
2 This range is underneath and above the operating temperature range, within the calibration load is fully functional
and could be used without damage.
Rosenberger Hochfrequenztechnik GmbH & Co. KG
P.O.Box 1260 D-84526 Tittmoning Germany
www.rosenberger.de
Tel. : +49 8684 18-0
Email : info@rosenberger.de