03S12L-000S3 Datasheet, PDF (2/3 Pages) Rosenberger Hochfrequenztechnik GmbH & Co. KG – Open Circuit Plug
Technical Data Sheet
RPC-3.50
Open Circuit
Plug
03S12L-000S3
Electrical data
Frequency range
Return loss
DC to 26.5 GHz
≤ 0.10 dB, DC to 4 GHz
≤ 0.12 dB, 4 GHz to 8 GHz
≤ 0.18 dB, 8 GHz to 26.5 GHz
Error from nominal phase1
≤ 1.0°, DC to 4 GHz
≤ 1.5°, 4 GHz to 8 GHz
≤ 2.0°, 8 GHz to 26.5 GHz
1 The nominal phase is defined by the Offset Delay, the Offset Loss and the Fringing Capacitances.
Mechanical data
Mating cycles
Maximum torque
Recommended torque
Gauge
≥ 500
1.70 Nm
0.90 Nm
0.00 mm to 0.04 mm
General standard definitions
For proper operation the vector network analyzer (VNA) needs a model describing the electrical behaviour
of this calibration standard. The different models, units, and terms used will depend on the VNA type and they
will have to be entered into the VNA. All values are based on typical geometry and plating.
Offset Zo / Impedance / Zo
Offset Delay
Length (electrical) / Offset Length
Offset Loss
Loss
Fringing Capacitances2
50 Ω
23.3495 ps
7.00 mm
2.20 GΩ/s
0.0089 dB/ GHz
2 Fringing Capacitances are determined individually for each open circuit and are documented in a Calibration Certificate.
Environmental data
Operating temperature range3
Rated temperature range of use4
Storage temperature range
+ 20 °C to +26 °C
0 °C to +50 °C
- 40 °C to +85 °C
RoHS
compliant
3 Temperature range over which these specification are valid.
4 This range is underneath and above the operating temperature range, within the open circuit is fully functional
and could be used without damage.