English
Language : 

02S119-S20D3 Datasheet, PDF (2/3 Pages) Rosenberger Hochfrequenztechnik GmbH & Co. KG – Calibration Adaptor
Technical Data Sheet
Calibration Adaptor
RPC-2.92 Plug – SMP Plug full detent
02S119-S20D3
Electrical data
Frequency
Return loss
DC to 40 GHz
≥ 32 dB, DC to 12 GHz
≥ 26 dB, 12 GHz to 26.5 GHz
≥ 21 dB, 26.5 GHz to 40 GHz
Mechanical data
Mating cycles
Maximum torque
Recommended torque
Engagement force
Disengagement force
Gauge
RPC-2.92
≥ 500
1.70 Nm
0.90 Nm
0.00 mm to 0.03 mm
SMP
≥ 100
Full detent 68 N
Full detent 22 N
0.00 mm to 0.05 mm
General standard definition
For proper operation the vector network analyzer (VNA) needs a model describing the electrical behaviour
of this calibration standard. The different models, units, and terms used will depend on the VNA type and they
will have to be entered into the VNA. All values are based on typical geometry and plating.
Offset Zo / Impedance / Zo
Offset Delay
Length (electrical) / Offset Length
Offset Loss
Loss
50 Ω
51.402 ps
15.41 mm
4.00 GΩ/s
0.0179 dB/ GHz
Environmental data
Operating temperature range1
Rated temperature range of use2
Storage temperature range
+20 °C to +26 °C
0 °C to +50 °C
-40 °C to +85 °C
RoHS
compliant
1 Temperature range over which these specifications are valid.
2 This range is underneath and above the operating temperature range, within the open circuit is fully functional
and could be used without damage.
Rosenberger Hochfrequenztechnik GmbH & Co. KG
P.O.Box 1260 D-84526 Tittmoning Germany
www.rosenberger.de
Tel. : +49 8684 18-0
Email : info@rosenberger.de
Page