English
Language : 

AB081C40W Datasheet, PDF (3/3 Pages) Roithner LaserTechnik GmbH – High Power Single Bar CW Infrared Laser Diode
Notes
1. High power laser diode arrays are high energy laser devices. It is harmful to human body
and health. Never look directly into the laser output port!
2. The storage temperature is between –40 and 85 °C.
3. In generally, the lifetime can be shortened by high temperature. So the generated heat
must be removed in time when the LD is working. The water cooling system, forced air
cooling or TEC system are recommended to keeping the LD at a suitable temperature.
4. Constant-current power supply with voltage regulator should be used to avoid surge. To
use a laser diode in following sequences: Turn on the power supply; connect to the laser
diode; increase the current gradually, and then turn off the power. Please make sure that
the power supply has no current overshoot at any time. The current overshoot can
damage the laser diodes permanently.
5. The high power laser diode arrays are very sensitive to electrostatic. Please wear
anti-static bracelet during operating with the laser diode arrays.
6. Be sure that the operating current does not exceed the specified operating current.
Otherwise it will accelerate laser aging, shorten lifetime or even damage devices
permanently.
7. A clean, dry and ventilated environment should be available when storing and operating
laser diode arrays. Dust may degrade the laser diode arrays.
20.09.2010
AB081C40W
3 of 3