English
Language : 

REM6-053.3S Datasheet, PDF (4/6 Pages) Recom International Power – 6 Watt 2:1 & 4:1 DIP24 Single and Dual Output
REM6
DC/DC Converter
Specifications (measured at TA= 25°C, nominal input voltage, full load and after warm-up)
Series
Isolation Voltage
Means of Protection
Leakage Current
Medical Device Classification
Internal Clearance
Creepage
External Clearance
Creepage
Isolation Capacitance
Insulation Grade
I/P to O/P
working voltage
240VAC, 60Hz
I/P to O/P
I/P to O/P
I/P to O/P,
I/P to O/P,
I/P to O/P,
I/P to O/P,
Single Output
Dual Output
Single Output
Dual Output
Notes:
Note5: This Power module is not internally fused. A input line fuse must be always used.
5kVAC / 1 minute
250VAC / continuous
2MOPP
2µA
Type CF applied device (design to meet)
8mm
8mm
>19.72mm
>14.64mm
>19.72mm
>14.64mm
12pF typ. / 17pF max.
Reinforced Insulation
Recomended Fuse:
2:1 Input Voltage
5V
12V
24V
48V
Fuse (slow blow)
T2.5A
T1.25A
T0.63A
T0.315A
4:1 Input Voltage
24V
48V
Fuse (slow blow)
T1.25A
T0.63A
ENVIRONMENTAL
Parameter
Relative Humidity
Temperature Coefficient
Thermal Impedance
max. Case Temperature Range
max. Ambient Temperature Range
MTBF (+25°C)
SAFETY AND CERTIFICATIONS
Certificate Type (Safety)
CB Medical Safety
ANSI/AAMI
CAN/CSA Medical
Certificate Type (Others)
EMI Standard (7)
ESD
Radiated Immunity
Fast Transient (6)
Surge (6)
Conducted Immunity
MED-4
Condition
natural convection (20LFM)
according to MIL-HDBK-217F, full load
Report / File Number
E314885-A6
1409015
E314885-A6
E314885-A6
Conditions
Conducted
Radiated
Conducted and Radiated
Air ±8kV; Contact ±6kV
10V/m
±2kV
±2kV
20Vr.m.s
continued on next page
REV.: 05/2016
Value
5% to 95% RH
±0.02% / °C
18°C / W
-40°C to +105°C
see calculation example
4718 x 103 hours
Standard
IEC-60601-1
Medical Report + ISO14971 Risk Assessment
ES60601-1
C22.2 No. 60601-1:08
Standard / Criterion
EN55011 (EN-55022), Class A, B
EN55011 (EN-55022), Class A, B
FCC18
EN61000-4-2, Criteria A
EN61000-4-3, Criteria A
EN61000-4-4, Criteria A
EN61000-4-5, Criteria A
EN61000-4-6, Criteria A