English
Language : 

REM6 Datasheet, PDF (1/5 Pages) Recom International Power – Reinforced Insulation for 250VAC Working Voltage
Features
Regulated
Converters
• Reinforced Insulation for 250VAC Working Voltage
• Clearance and Creepage Distance: 8mm
• 5kVAC I/P to O/P 2MOPP Isolation
• 2µA Patient Leakage Current
• Industry Standard Pinout
• 2:1 and 4:1 Wide Input Range
DC/DC Converter
REM6
Description
The REM6 series of medical grade regulated DC/DC converters features reinforced 5kVAC/1 minute
isolation with low 2µA leakage and are 60601-1 3rd Ed. certified for 250VAC continuous working. The
compact DIP24 package offers tightly regulated single and dual outputs, even under no-load conditions.
The outputs are short circuit and overload protected. The converters are available in two different pin-
ning options and optionally with an external control pin for standby consumption as low as 12.5mW. The
converters are fully certified to CB, IEC/EN and ANSI/AAMI standards and carry the UL mark.
6 Watt
2:1 & 4:1
55 YEAR
warranty
DIP24
Single and Dual
Selection Guide
Output
Part
Number
REM6-xx3.3S/ (3,4)
REM6-xx05S/ (3,4)
REM6-xx12S/ (3,4)
REM6-xx15S/ (3,4)
REM6-xx24S/ (3,4)
REM6-xxx05D/ (3,4)
REM6-xx12D/ (3,4)
REM6-xx15D/ (3,4)
REM6-xx3.3SW/ (3,4)
REM6-xx05SW/ (3,4)
REM6-xx12SW/ (3,4)
REM6-xx15SW/ (3,4)
REM6-xx24SW/ (3,4)
REM6-xx05DW/ (3,4)
REM6-xx12DW/ (3,4)
REM6-xx15DW/ (3,4)
nom. Input
Voltage (1)
[VDC]
5 / 12 / 24 / 48
5 / 12 / 24 / 48
5 / 12 / 24 / 48
5 / 12 / 24 / 48
5 / 12 / 24 / 48
5 / 12 / 24 / 48
5 / 12 / 24 / 48
5 / 12 / 24 / 48
24 / 48
24 / 48
24 / 48
24 / 48
24 / 48
24 / 48
24 / 48
24 / 48
Output
Voltage
[VDC]
3.3
5
12
15
24
±5
±12
±15
3.3
5
12
15
24
±5
±12
±15
Output
Current
[mA]
Efficiency
typ.
[%]
Max. Capacitive
Load
[µF]
1800 81.5 / 83.5 / 83 / 82.5 2100
1200 86 / 86 / 86 / 86.5 1500
500
86 / 89 / 89 / 88
260
400 87.5 / 88.5 / 88.5 / 88.5 210
250 87 / 88.5 / 88.5 / 88 75
±600 84 / 85 / 85 / 85 ±860
±250 86.5 / 89 / 88.5 / 88 ±150
±200 87.5 / 88 / 88.5 / 87 ±110
1800
83 / 82.5
2100
1200
86 / 86.5
1500
500
89 / 88
260
400
89 / 88.5
210
250
88.5 / 88
75
±600
85 / 85
±860
±250
88.5 / 88
±150
±200
88.5 / 87
±110
E314885
2MOPP
250VAC
Model Numbering
REM 6- __ __ __ __ / __ __
Output Power
Input Voltage (1)
Output Voltage
Single or Dual
CTRL Option (4)
Pinning (3)
Input Type (2)
IEC-60601-1 Certified
ES-60601-1 Certifed
EN-55011 Certified
EN-55022 Certified
Notes:
Note1: for 4:1 Input Voltage Type add “W”, see Note 2.
2:1
nom. Vin
4:1 “W”
xx= 4.5-9 Vin = “05”
xx= 9-36Vin
xx= 9-18Vin = “12”
xx= 18-75Vin
xx= 18-36Vin = “24”
xx= 36-75Vin = “48”
nom. Vin
= “24”
= “48”
Note2: Blank for Standard 2:1 Input Voltage Range; „W“ suffix for 4:1 Input Voltage Range
Note3: „A“ suffix for A pinning; „C“ suffix for C pinning, for more details refer to Package Style and Pinning
Note4: „CTRL“ suffix for control pin option, for A pinning only, for C pinning not aviable
Examples:
REM6-0512D/A
=
REM6-1215S/C
=
REM6-4815SW/A/CTRL =
REM6 -243.3SW/C
=
2:1 Input,
2:1 Input,
4:1 Input,
4:1 Input,
4.5-9Vin,
9-18Vin,
36-75Vin,
9-36Vin,
±12Vout,
15Vout,
15Vout,
3.3Vout,
pinout „A“, without control pin
pinout „C“, without control pin
pinout „A“ with control pin
pinout „C“, without control pin
www.recom-power.com
REV.: 06/2016
MED-1