English
Language : 

PM66 Datasheet, PDF (1/2 Pages) Protek Devices – 51-65 WATT MEDICAL SWITCHING POWER SUPPLIES
PROTEK 51-65 WATT MEDICAL SWITCHING POWER SUPPLIES
DESCRIPTION
The PM66 series of compact, open PCB constructed,
AC-DC switching power supplies are capable of delivering
51 to 65 watts of continuous power. They operate at 90 to
264VAC input voltage without the need of voltage
selection. They are ideally suited for use in medical
equipment not for life support. All models meet the safety
requirements of UL, CSA and IEC.
PM66 SERIES
( LV D )
RoHS
FE AT U R E S
• Recognized or certified by UL,CSA and TÜV
• Small size
• 100% burn-in
• Wide input range 90 to 264VAC
• Input surge current protection
• Overvoltage protection
• Overcurrent protection
• Compliant with RoHS requirements
Safety Standard Approvals:
INPUT SPECIFICATIONS
Input voltage:
90 to 264VAC
Input frequency:
47 to 63 Hz
Input current:
1.60A( rms) for 115VAC
1.00A( rms) for 230VAC
Earth le aka ge cu rren t: 90 uA m ax. @ 115VAC, 6 0H z
150uA max. @ 230VAC, 50Hz
UL 2601-1, CSA C22.2 NO. 601.1
File No. E178020
TÜV EN60601-1
OUTPUT SPECIFICATIONS
Output voltage/current: See rating chart
Total output power:
65 watts maximum
Ripple and noise:
1﹪peak to peak maximum
Overvoltage protection: Provided on output; set at
112 - 132﹪ of its nominal
output voltage
Overcurrent protection: The output protected to short
circuit conditions
Temperature coefficient: All outputs ±0.04﹪ /℃
maximum
Transient response:
Maximum excursion of 4﹪
or better on all models,
recovering to 1﹪of final value
within 500us after a 25﹪
step load change
EN V IR O NM ENTAL S P EC IF IC AT IO N S
Operating temperature: 0℃to +70℃
Storage temperature: -40℃to +85℃
R e l a t i v e h u m i d i t y:
5﹪ to 95﹪ non-condensing
Derating:
Derate from 100﹪at +50℃
linearly to 50﹪at +70℃
2-11
GENERAL SPECIFICATIONS
Switching frequency: 42KHz ±5KHz
E f f i c i e n c y:
75﹪minimum on models with
Vo≧12V, 66% minimum on the others
Hold-up time:
10 msec minimum at 110VAC
Line regulation:
Inrush current:
Withstand voltage:
MTBF:
±0.5﹪maximum at full load
17 amps @ 115VAC or 40 amps
@230VAC, at 25℃ cold start
4000VAC from input to output
1500VAC from input to ground
500VAC from output to ground
400,000 hours minimum at full load
at 25℃ ambient , calculated per
MIL-HDBK-217F
EMC Performance (EN60601-1-2: 2001)
EN55011:
Class B conducted, Class B radiated
FCC:
Class B conducted, Class B radiated
VCCI:
Class B conducted, Class B radiated
EN61000-3-2:
Harmonic distortion, Class A
EN61000-3-3:
Line flicker
EN61000-4-2:
ESD, ±8KV air and ±6KV contact
EN61000-4-3:
EN61000-4-4:
Radiated immunity, 3V/m for
80~2500MHz
Fast transient/burst, ±2KV
EN61000-4-5:
Surge, ±1KV diff., ±2KV com.
EN61000-4-6:
Conducted immunity, 3Vrms
EN61000-4-8:
Magnetic field immunity, 3A/m
E N 6 1 0 0 0 - 4 - 11 :
Voltage dips, 30% reduction for 500ms,
60% reduction for 100ms and >95%
reduction for 10ms