PM66 Datasheet, PDF (1/2 Pages) Protek Devices – 51-65 WATT MEDICAL SWITCHING POWER SUPPLIES | |||
|
PROTEK 51-65 WATT MEDICAL SWITCHING POWER SUPPLIES
DESCRIPTION
The PM66 series of compact, open PCB constructed,
AC-DC switching power supplies are capable of delivering
51 to 65 watts of continuous power. They operate at 90 to
264VAC input voltage without the need of voltage
selection. They are ideally suited for use in medical
equipment not for life support. All models meet the safety
requirements of UL, CSA and IEC.
PM66 SERIES
( LV D )
RoHS
FE AT U R E S
⢠Recognized or certified by UL,CSA and TÃV
⢠Small size
⢠100% burn-in
⢠Wide input range 90 to 264VAC
⢠Input surge current protection
⢠Overvoltage protection
⢠Overcurrent protection
⢠Compliant with RoHS requirements
Safety Standard Approvalsï¼
INPUT SPECIFICATIONS
Input voltageï¼
90 to 264VAC
Input frequencyï¼
47 to 63 Hz
Input currentï¼
1.60Aï¼ rmsï¼ for 115VAC
1.00Aï¼ rmsï¼ for 230VAC
Earth le aka ge cu rren tï¼ 90 uA m ax. @ 115VAC, 6 0H z
150uA max. @ 230VAC, 50Hz
UL 2601-1, CSA C22.2 NO. 601.1
File No. E178020
TÃV EN60601-1
OUTPUT SPECIFICATIONS
Output voltage/currentï¼ See rating chart
Total output powerï¼
65 watts maximum
Ripple and noiseï¼
1﹪peak to peak maximum
Overvoltage protectionï¼ Provided on output; set at
112 - 132﹪ of its nominal
output voltage
Overcurrent protectionï¼ The output protected to short
circuit conditions
Temperature coefficientï¼ All outputs ±0.04﹪ /â
maximum
Transient responseï¼
Maximum excursion of 4﹪
or better on all models,
recovering to 1﹪of final value
within 500us after a 25﹪
step load change
EN V IR O NM ENTAL S P EC IF IC AT IO N S
Operating temperature: 0âto +70â
Storage temperatureï¼ -40âto +85â
R e l a t i v e h u m i d i t yï¼
5﹪ to 95﹪ non-condensing
Deratingï¼
Derate from 100﹪at +50â
linearly to 50﹪at +70â
2-11
GENERAL SPECIFICATIONS
Switching frequencyï¼ 42KHz ±5KHz
E f f i c i e n c yï¼
75﹪minimum on models with
Voâ§12V, 66% minimum on the others
Hold-up timeï¼
10 msec minimum at 110VAC
Line regulationï¼
Inrush currentï¼
Withstand voltageï¼
MTBFï¼
±0.5﹪maximum at full load
17 amps @ 115VAC or 40 amps
@230VAC, at 25â cold start
4000VAC from input to output
1500VAC from input to ground
500VAC from output to ground
400,000 hours minimum at full load
at 25â ambient , calculated per
MIL-HDBK-217F
EMC Performance (EN60601-1-2: 2001)
EN55011:
Class B conducted, Class B radiated
FCC:
Class B conducted, Class B radiated
VCCI:
Class B conducted, Class B radiated
EN61000-3-2:
Harmonic distortion, Class A
EN61000-3-3:
Line flicker
EN61000-4-2:
ESD, ±8KV air and ±6KV contact
EN61000-4-3:
EN61000-4-4:
Radiated immunity, 3V/m for
80~2500MHz
Fast transient/burst, ±2KV
EN61000-4-5:
Surge, ±1KV diff., ±2KV com.
EN61000-4-6:
Conducted immunity, 3Vrms
EN61000-4-8:
Magnetic field immunity, 3A/m
E N 6 1 0 0 0 - 4 - 11 :
Voltage dips, 30% reduction for 500ms,
60% reduction for 100ms and >95%
reduction for 10ms
|
▷ |