English
Language : 

TR30M Datasheet, PDF (1/1 Pages) Powerbox – 20 - 30 WATTS SINGLE OUTPUT AC/DC MEDICAL
TR30M Series
20-30 WATTS SINGLE OUTPUT AC/DC MEDICAL
Features
• Wide input voltage 90VAC to 264VAC, 50/60Hz.
• Continuous short circuit protection.
• Over voltage protection.
• Output voltage available from 5VDC to 24VDC.
• High efficiency 75% typical.
• Meets EN60601-1, EN55022 Class ”B”.
Specifications
INPUT
Voltage Range
90 to 264VAC
Frequency
47Hz to 63Hz
Inrush current
50A max @ 264VAC.
Input current
0.6A.
Isolation
Input to output 5,656VDC.
OUTPUT
Output Voltage
Factory preset, see table (not adjustable)
Voltage Set-Point
+/-2% to +/-4% (60% load)
Ripple and Noise
1% p-p max
Short Circuit Protection Continuous (Auto recovery)
Over Voltage Protection Yes.
Hold-up Time
10ms typ @ 115Vac
Load Regulation
+/-2 to +/-6% (20% to 100% load)
Line Regulation
+/-1% max at full load
ENVIRONMENTAL
Operating Temperature 0°C to +40°C
Storage Temperature
-20°C to +85°C
Technical Illustration
STANDARDS AND APPROVALS
Safety standards
EN60601-1
C-Tick
AS/NZS CISPR11 Group 1, Class A
EMI standards
CISPR/FCC Class B
MEPS compliance
TBC
MECHANICAL
Dimensions
110x50x30mm
Weight
220g
AC Inlet
IEC320/C8.
Output Connector
Right angle, hollow coaxial plug, 2.5mm inside
diameter; 5.5mm outside diameter; 9.5mm length
Center positive polarity.
DC Cord Length
720mm with ferrite core 18AWG/UL1185.
Selection Table
MODEL
NUMBER
RATED
Vnom
TR30M050-21A11
5Vdc
TR30M090-21A11
9Vdc
TR30M120-21A11
12Vdc
TR30M150-21A11
15Vdc
TR30M180-21A11
18Vdc
TR30M240-21A11
24Vdc
OUTPUT
CURRENT
4.0A
3.0A
2.5A
2.0A
1.65A
1.25A
MAX POWER
20W
27W
30W
30W
30W
30W
Your dependable power partner – www.powerbox.com.au