PMP30 Datasheet, PDF (1/2 Pages) Powerbox – 25 - 30 WATTS SINGLE OUTPUT AC/DC MEDICAL
PMP30 Series
25-30 WATTS SINGLE OUTPUT AC/DC MEDICAL
Features
• Input surge current protection.
• Overvoltage protection.
• Overcurrent protection.
• Short-circuit protection with auto recovery.
• 100% burn-in at full rated load.
Specifications
INPUT
Input Voltage
Frequency
Input Current
Leakage current
Isolation
Inrush Current
OUTPUT
Output Voltage
Output Current
Ripple & Noise
Line Regulation
Load Regulation
Transient Response
Overvoltage Protection
Overcurrent Protection
Temperature Coefficient
Holdup Time
90-264Vac
47Hz–67Hz
0.80A (rms) for 115Vac.
0.50A (rms) for 230Vac.
90µA max. @ 115Vac. 60Hz.
150µA max. @ 230Vac. 50Hz.
Input – Output: 4000Vac
Input – Ground: 1500Vac
Output – Ground: 500Vac
15A @ 115V, 30A @ 230V
See table
See table
1% peak to peak
±0.5% maximum at full load.
See table
Max. deviation 4% recovering to 1% within 500µs
after 25% load charge
Provided on V1 only. Set at 110%–132% of
nominal.
All outputs protected for short circuit conditions
All outputs ±0.04% per °C.
Typically 20ms at 115Vac.
OPERATING
MTBF
400,000 hrs @ 25°C
Efficiency
65-70% minimum.
ENVIRONMENTAL
Operating Temperature
0°C to 70°C
Derating
Derate to 50% at 70°C
Storage Temperature
-40°C to 85°C
Relative Humidity
5%–95% non-condensing
STANDARDS AND APPROVALS
Safety standards
UL2601-1, CSA C22.2, TUV EN60601-1
C-Tick
AS/NZS CISPR11 Group 1, Class A
EMC standards
EN60601-1-2: 2001,EN61000-3-2, -3,
EN61000-4-2, -3, -4, -5, -6, -8, -11
EMI standards
EN55011 Class B, FCC Class B, VCCI Class B
MEPS compliance
TBC
MECHANICAL
Dimensions
147x75x42mm
Weight
0.55kgs approx.
Technical Illustration
Your dependable power partner – www.powerbox.com.au
359