English
Language : 

PM40AL Datasheet, PDF (1/2 Pages) Tumbler Technologies + TRUMPower – 40 WATT MEDICAL POWER SUPPLIES
PM40AL Series
40 WATTS SINGLE & MULTIPLE OUTPUT AC/DC MEDICAL
Features
• Low safety ground leakage current.
• Wide input range 85 to 264Vac.
• Input surge current protection.
• Overvoltage protection.
• Overcurrent protection.
• Open PCB construction.
• 100% burn-in at full rated load.
Specifications
INPUT
Input Voltage
85–264Vac wide input
Frequency
47Hz–63Hz
Input Current
Leakage current
Isolation
1.2A (rms) for 115Vac.
0.7A (rms) for 230Vac.
90µA max. @ 115Vac. 60Hz.
150µA max. @ 230Vac. 50Hz.
Input – Output: 4000VAC
Input – Ground: 1500VAC
Output – Ground: 500VAC
Inrush Current
15A @ 115V, 30A @ 230V
OUTPUT
Output Voltage
See table
Output Voltage Adjustment V1 ±10%
Output Current
See table
Ripple & Noise
1% peak to peak
Line Regulation
±0.5% maximum at full load.
Load Regulation
See table
Transient Response
Overvoltage Protection
Max. deviation 4% recovering to 1% within
500µs after 25% load charge
Provided on V1 only. Set at 110%–132% of
nominal
Overcurrent Protection
All outputs protected for short circuit conditions
Temperature Coefficient All outputs ±0.04% per °C
Holdup Time
Typically 10ms at 110VAC
OPERATING
MTBF
600,000 hrs @ 25°C
Efficiency
65%–70%
ENVIRONMENTAL
Operating Temperature
0°C to 70°C
Derating
Derate to 50% at 70°C
Storage Temperature
-40°C to 85°C
Relative Humidity
5%–95% non-condensing
STANDARDS AND APPROVALS
Safety standards
UL2601-1, CSA C22.2, TUV EN60601-1
C-Tick
AS/NZS CISPR11 Group 1, Class A
EMC standards
EMI standards
EN60601-1-2: 2001,EN61000-3-2, -3,
EN61000-4-2, -3, -4, -5, -6, -8, -11
EN55011 Class B, FCC Class B, VCCI Class B
MECHANICAL
Dimensions
127x76.2x30mm
Weight
230grams.
334