English
Language : 

PM200 Datasheet, PDF (1/2 Pages) Extech Instruments Corporation. – Fiber Optic Power Meter
PM200 Series
200 WATTS SINGLE & MULTIPLE OUTPUT AC/DC MEDICAL
Features
• Low safety ground leakage current.
• Autoranging input.
• Overvoltage protection.
• Overcurrent protection.
• Power fail detection (PFD) signal.
• Optional cover and fan assembly.
• 100% burn-in at full rated load.
• Short curcuit protected with auto-recvery
Specifications
INPUT
Input Voltage
90-130/180-264Vac (Autoranging)
Frequency
47Hz–67Hz
Input Current
Leakage current
Isolation
4.5A (rms) for 115Vac.
2.6A (rms) for 230Vac.
90µA max. @ 115Vac. 60Hz.
150µA max. @ 230Vac. 50Hz.
Input – Output: 4000VAC
Input – Ground: 1500VAC
Output – Ground: 500VAC
Inrush Current
20A @ 115V, 40A @ 230V
OUTPUT
Output Voltage
See table
Output Voltage Adjustment V1 ±10%
Output Current
See table
Ripple & Noise
2% peak to peak
Line Regulation
±0.5% maximum at full load.
Load Regulation
See table
Transient Response
Overvoltage Protection
Max. deviation 4% recovering to 1% within 500µs
after 25% load charge
Provided on V1 only. Set at 110%–132% of
nominal
Overcurrent Protection
All outputs protected for short circuit conditions
Temperature Coefficient All outputs ±0.04% per °C
Holdup Time
Typically 12mS at 110VAC
PFD Signal
Power fail detection signal. TTL logic low upon
loss of input power.
OPERATING
MTBF
450,000 hrs @ 25°C
Efficiency
70% minimum.
ENVIRONMENTAL
Operating Temperature
0°C to 70°C
Derating
Derate to 50% at 70°C
Storage Temperature
-40°C to 85°C
Relative Humidity
5%–95% non-condensing
Cooling
25 CFM forced air, provided on “C” version, to be
provided for “B”version by user.
STANDARDS AND APPROVALS
Safety standards
UL2601-1, CSA C22.2, TUV EN60601-1
C-Tick
AS/NZS CISPR11 Group 1, Class A
EMC standards
EN60601-1-2: 2001,EN61000-3-2, -3,
EN61000-4-2, -3, -4, -5, -6, -8, -11
EMI standards
EN55011 Class B, FCC Class B, VCCI Class B
MECHANICAL
Dimensions
229x127x65mm
Weight
1.4Kgs for “B” version.
1.6Kgs for “C” version.