English
Language : 

OBP02013 Datasheet, PDF (1/3 Pages) Powerbox – 200-300 WATTS SINGLE & MULTIPLE OUTPUT AC/DC MEDICAL
Medline 300 Series
200-300 WATTS SINGLE & MULTIPLE OUTPUT AC/DC MEDICAL
Features
• EN61000-3-2 class A and D compliant.
• Power factor 0.98 typical.
• Remote sense on output 1 and output 2.
• Overvoltage protection.
• Overcurrent protection.
• Power fail detection (PFD) signal.
• 100% burn-in at full rated load.
• Remote inhibit - TTL high disables output.
• Short curcuit protection
Specifications
INPUT
Input Voltage
Frequency
Input Current
Leakage current
Isolation
Inrush Current
OUTPUT
Power
Voltage
Ripple & Noise
Overvoltage Protection
Overcurrent Protection
Temperature Coefficient
Transient Response
PFD Signal
Remote inhibit
Line Regulation
Holdup Time
90-264Vac
47Hz–63Hz
4.7A (rms) for 115Vac
2.2A (rms) for 230Vac
90µA max. at 115Vac. 60Hz
150µA max. at 230Vac. 50Hz
Input – Output: 4000VAC
Input – Ground: 1500VAC
Output – Ground: 500VAC
30A at 115V or 80A at 230V at 25°C and cold
start
See table
See table
2% peak to peak max
Provided on output #1 only. Set at 115%–140%
of nominal output voltage
All outputs protected for short circuit conditions
All outputs ±0.04% per °C
Max. excursion of 4%or better on all models,
recovering to 1% of final value within 500µs after
a 25% step load change
TTL logic high for normal operation and TTL logic low
upon loss of input power. This signal appears at least
1ms prior to +5V output dropping 5% below its
nominal value. This signal also provides a minimum
delay of 100ms after +5V is within regulation
Requires an external TTL high level signal to
inhibit outputs for standard models
±0.2% maximum at full load
Typically 12ms at 110VAC
OPERATING
MTBF
350,000 hrs @ 25°C
Efficiency
70% minimum.
ENVIRONMENTAL
Operating Temperature
Derating
Storage Temperature
0°C to 70°C
Derate from 100% at +50°C linearly to 50%
at 70°C
-40°C to 85°C
Humidity
5%–95% non-condensing
Cooling
35 CFM forced air, provided on “C” version, to be
provided for “B” version by user
STANDARDS AND APPROVALS
Safety standards
UL60601-1, TUV EN60601-1
C-Tick
AS/NZS CISPR11 Group 1, Class A
EMC standards
MECHANICAL
EN55011 Class B, FCC Class B, EN61000-3-2, -3,
EN61000-4-2, -3, -4, -5, -6, -8, -11
Dimensions
Weight
177.8x101.6x38.1mm
0.82Kgs for “B” version.
0.96Kgs for “C” version.
Your dependable power partner – www.powerbox.com.au