PE300BF and PE300BUV
Operational Specifications
Description
Nominal
Range
Power
300 watts
180-320 watts
Current
21 amps (DC)
10-22 amps (DC)
Operating Voltage 14 volts (DC)
13-16 volts (DC)
Ignition Voltage
23 kilovolts (recommended minimum)
Temperature
150 Þ C (maximum)
Lifetime*
1000 hours (500 hour warranty)
* End of lamp life is defined as 50% of initial output.
Output at Nominal Power
F= UV Filtered Output/ UV = UV Enhanced Output
Description
PE300BF
Peak Intensity
515x103 candelas
Radiant Output*
50 watts
UV Output*
2.6 watts
IR Output*
28.8 watts
Visible Output*
5000 Lumens
Color Temperature 5600 Kelvin
Peak Instabilities 4%
Beam Geometry** 5º/6º/7º
*These values indicate total output in all directions.
PE300BUV
460x103 candelas
50 watts
6.6 watts
26.8 watts
4500 Lumens
5050 Kelvin
4%
5º/6º/7º
Wavelengths = UV<390nm, IR>770nm, Visible 390nm to 770nm.
** Beam Geometry defined as half angle at 10% PTS after 01/100/1000 hours.
Dimensions
.165
(4.19)
1.000
(25.4)
.250 (6.35)
.250 (6.35)
1.000
(25.4)
6-32 UNC 2B X .156 DP
.125 (3.18)
2 x .40
(10.2)
1.667
(42.34)
.035
(.899)
.066 (1.52)
1.298 (32.96)
Y
X
Focal Point
Cathode
Dimensions in parentheses are in millimeters.
Physical Specifications
Description
Arc Gap
Reflector Geometry
Weight
Window Diameter
Specification
.049 inch (1.24mm)
Parabolic Y2 = 0.5 X (inch)
132 grams
1 inch (25.4 mm)
Focused Output at F/1.0 Lens
Description
Visible Output
6mm aperture
3130 Lumens
8mm aperture
1410 Lumens
* Nominal values at 300 watts after 2 hour burn-in.
Total Output*
29 watts
14 watts
Notes
1. Lamp must not be operated with window facing upwards within
45º of vertical.
2. Seal temperature must not exceed 150º C.
3. Current/power regulated power supplies and PerkinElmer lamp
housing units are recommended.
4. Lamp must be operated within recommended current and power
range. Over powering may lead to arc instability, hard starting
and premature aging.
5. Hot mirror assembly is available for IR filtering.
6. CERMAX lamps are much safer lamps to use than their quartz
xenon arc lamp equivalents. However, caution must be practiced
when operating lamps because they are under high pressure,
require high voltage, reach temperatures up to 200º C, and their
IR and UV radiation can cause skin burns and eye damage.
Read hazard sheet included with each lamp shipment.
Spectral Output
F
UV
100
80
60
40
F + HOT MIRROR
20
0
200
300 400 500
600 700
800
900 1000 1100
WAVELENGTH (NM)
PerkinElmer welcomes inquiries about special lamps. We would be pleased to discuss the requirements of your application and the feasibility of
designing a lamp specifically suited to your needs.
For more information e-mail us at opto@perkinelmer.com or visit our web site at www.perkinelmer.com/opto.
All values are nominal; specifications subject to change without notice.
USA:
Europe:
Asia:
PerkinElmer Optoelectronics
PerkinElmer Optoelectronics GmbH
PerkinElmer Optoelectronics
44370 Christy St.
Wenzel-Jaksch-Str. 31
47 Ayer Rajah Crescent #06-12
Fremont, CA 94538
65199 Wiesbaden
Singapore 139947
Phone: (510) 979-6500
Germany
Phone: +65 775 2022
Fax: (510) 687-1140
Phone: +49 611 492 0
Fax: +65 775 1008
.
Toll Free: (800) 775-6786
Fax: +49 611 492 369
© 2000 PerkinElmer inc. All rights reserved.
www.perkinelmer.com/opto
DS-156 Rev C 1101