English
Language : 

NLSL06S01A-E Datasheet, PDF (3/9 Pages) NICHIA CORPORATION – SPECIFICATIONS FOR WARM WHITE STANDARD LED
DIRECTIONS FOR USE
NICHIA STS-DA1-2841C <Cat.No.131112>
(1) Auxiliary Equipment and System Requirements
● The following figure shows how the Luminous Flux Standard LED is connected.
This product
Peltier Controller
Constant Current Power
Supply for LED
Digital Multimeter
for Forward Voltage
Measurements
Recommended specifications for the peripheral equipment
Peltier Controller
Constant Current Power Supply for LED
Digital Multimeter for Forward Voltage Measurements
PID Controller
Temperature measurement accuracy of ±0.03°C.
Drive current accuracy: ≤ 0.1% of the set current
Input impedance: ≥ 1 GΩ
(2) Installation
● The following figure shows how the product should be held by a holder.
(3) Measurement
● The product should be measured when luminous flux or radiant flux has stabilized after the start of
temperature control and constant current drive.
Typically, this product will be stable and ready to be used for calibration within 1 to 3 minutes.
Please refer to “STABILITY” on the following page for the characteristics during the first 10 minutes
after the LED starts emission.
2