English
Language : 

NI9203 Datasheet, PDF (6/8 Pages) National Instruments Corporation – 8 Ch, ±20 mA, 200 kS/s, 16-Bit C Series Analog Current Input Module
Power Requirements
Power consumption from chassis
Active mode
Sleep mode
Thermal dissipation (at 70 °C)
Active mode
Sleep mode
Physical Characteristics
399 mW max
5 mW max
1.22 W max
824 mW max
Screw-terminal wiring
Ferrules
Torque for screw terminals
Weight
Safety
12 to 24 AWG copper conductor wire with 10 mm (0.39 in.) of insulation
stripped from the end
0.25 mm2 to 2.5 mm2
0.5 to 0.6 N · m (4.4 to 5.3 lb · in.)
162 g (5.7 oz)
If you need to clean the module, wipe it with a dry towel.
Safety Voltages
Connect only voltages that are within the following limits.
Channel-to-COM
±30 VDC max
Isolation
Channel-to-channel
No isolation between channels
Channel-to-earth ground
Continuous
250 Vrms, Measurement Category II
Withstand
2,300 Vrms, verified by a 5 s withstand test
Measurement Category II is for measurements performed on circuits directly connected to the electrical distribution system. This category refers to local-level electrical
distribution, such as that provided by a standard wall outlet (e.g., 115 V for U.S. or 230 V for Europe). Examples of Measurement Category II are measurements performed on
household appliances, portable tools, and similar products.
Caution Do not connect the NI 9203 to signals or use for measurements within Measurement Categories III or IV.
Safety Standards
This product is designed to meet the requirements of the following standards of safety for electrical equipment for measurement, control, and laboratory use:
IEC 61010-1, EN 61010-1
UL 61010-1, CSA 61010-1
Note For UL and other safety certifications, refer to the product label or the Online Product Certification section.
Hazardous Locations
U.S. (UL)
Class I, Division 2, Groups A, B, C, D, T4; Class I, Zone 2, AEx nC IIC T4
Canada (C-UL)
Class I, Division 2, Groups A, B, C, D, T4; Class I, Zone 2, Ex nC IIC T4
Europe (DEMKO)
EEx nC IIC T4
Environmental
National Instruments C Series modules are intended for indoor use only but may be used outdoors if installed in a suitable enclosure. Refer to the manual for the chassis you are
using for more information about meeting these specifications.
Operating temperature
(IEC 60068-2-1, IEC 60068-2-2)
–40 to 70 °C
Storage temperature
(IEC 60068-2-1, IEC 60068-2-2)
–40 to 85 °C
Ingress protection
IP 40
Operating humidity
(IEC 60068-2-56)
10 to 90% RH, noncondensing
Storage humidity
6/8