PXI-4132 Datasheet, PDF (3/5 Pages) National Instruments Corporation – 100 V Precision Source Measure Unit (SMU)
100 V Precision Source Measure Unit (SMU)
SMU Specifications1
Specifications subject to change without notice. For complete specifications,
see the NI PXI-4132 Specifications at ni.com/manuals.
Voltage Programming Accuracy/Resolution
Range
±10 V
±100 V
Resolution
50 μV
500 μV
Accuracy ± (% of output + offset)
1-Year, 23 °C ± 5 °C
0.025% + 3.0 mV
0.025% + 10 mV
Peak-to-Peak Noise, typical
(0.1 to 10 Hz)
70 μV
300 μV
Current Programming Accuracy/Resolution2
Range
10 µA
100 µA
1 mA
10 mA
100 mA
Resolution
500 pA
5 nA
50 nA
500 nA
5 µA
Accuracy ± (% of output + offset)
1-Year, 23 °C ± 5 °C
0.034% + 2.0 nA
0.034% + 20 nA
0.034% + 0.2 µA
0.034% + 2.0 µA
0.034% + 20 µA
Peak-to-Peak Noise, typical
(0.1 to 10 Hz)
90 pA
900 pA
9 nA
90 nA
900 nA
Voltage Measurement Accuracy/Resolution
Range
±10 V
±100 V
Resolution
10 µV
100 µV
Accuracy ± (% of output + offset)
1-Year, 23 °C ± 5 °C
0.02% + 2.0 mV
0.02% + 5.0 mV
1Temperature coefficient (Tempco) is 15% of accuracy specification per °C.
2Minimum programmable current limit/level is 2% of range.
Current Measurement Accuracy/Resolution
Range
10 µA
100 µA
1 mA
10 mA
100 mA
Resolution
10 pA
100 pA
1 nA
10 nA
100 nA
Accuracy ± (% of output + offset)
1-Year, 23 °C ± 5 °C
0.028% + 1.0 nA
0.028% + 10 nA
0.028% + 0.1 µA
0.028% + 1.0 µA
0.020% + 10 µA
Measurement Speed1
Maximum Operation Rates Per Second for 60 Hz (50 Hz)
ADC Aperture Time
1/64 PLC
1/8 PLC
1 PLC
Measure to Host
3490 (2900)
470 (390)
59.9 (49.9)
Source-Measure to Host2
1900 (1700)
425 (360)
59.0 (49.3)
Maximum source update rate............... 4,200 updates/s
Trigger in to source delay...................... 500 ns
Triggers
Input Triggers
Types......................................................
Sources..................................................
Polarity...................................................
Minimum pulse width............................
Destinations4. ........................................
Polarity...................................................
Pulse width............................................
Start, Source, Sequence
Advance, Measure
PXI Trigger lines 0–73
Configurable
100 ns
PXI Trigger lines 0–73
Active high (unconfigurable)
150 ns
Power Requirements
PXI power requirement......................... 10 W at 5 V, 1 W at 3.3 V, 2 W at 12 V
Environment
Maximum altitude ...............................
Pollution degree....................................
Indoor use only.
2,000 m
(at 25 °C ambient temperature)
2
1Does not include load-dependent settling time; niDCPower Auto Zero property/attribute set to Off.
2S ource-measure to host and maximum source update rate are performed with the source delay set to
200 µs. This is sufficient for the output to settle within 1 percent of the requested level with a simple
resistive load. As you adjust the source delay for your application’s requirements, maximum rates vary.
3Pulse widths and logic levels compliant with PXI specifications.
4Input triggers can be reexported.
BUY ONLINE at ni.com or CALL 800 813 3693 (U.S.)