778397-01 Datasheet, PDF (1/8 Pages) National Instruments Corporation – S Series Multifunction DAQ 12 or 16-Bit,1 to 10 MS/s, 4 Analog Inputs | |||
|
S Series Multifunction DAQ 12 or 16-Bit,
1 to 10 MS/s, 4 Analog Inputs
NI 6120, NI 6115, NI 6110, NI 6111
⢠2 or 4 analog inputs; dedicated
A/D converter per channel
⢠1 to 10 MS/s per channel
maximum sample rate
⢠Analog and digital triggering
⢠AC or DC coupling
⢠8 input ranges from ±200 mV
to ±42 V
⢠2 analog outputs at 4 MS/s single
channel or 2.5 MS/s dual channel
⢠8 digital I/O lines
(5 V TTL/CMOS)
⢠Two 24-bit counter/timers
⢠Measurement services that simplify
configuration and measurements
Operating Systems
⢠Windows 2000/NT/XP
⢠Mac OS X
⢠Linux
Recommended Software
⢠LabVIEW 7.x or higher
⢠LabWindows/CVI 7.x or higher
⢠Measurement Studio 7.x or higher
⢠Digital Waveform Editor
⢠SignalExpress 1.x or higher
Other Compatible Software
⢠VI Logger 2.x or higher
⢠Visual Studio .NET
⢠Visual Basic, C/C++, and C#
Measurement Services
Software (included)1
⢠NI-DAQmx driver software
⢠Measurement & Automation
Explorer configuration utility
⢠VI Logger Lite data-logging software
1Mac OS X and Linux applications
must use NI-DAQmx Base driver software.
Calibration Certificate Available
Analog
Input
Sampling
Input
Analog
Max Output
Family
Bus
Inputs Resolution (bits) Rate (MS/s) Range (V)
Outputs
Rate (MS/s)
NI 6120
PCI, PXI
4
16
13
±0.2 to ±42
2
41
NI 6115
PCI, PXI
4
12
10
±0.2 to ±42
2
41
NI 6110
PCI
4
12
5
±0.2 to ±42
2
41
NI 6111
PCI
2
12
5
±0.2 to ±42
2
41
14 MS/s single channel; 2.5 MS/s on two channels 2Hardware-timed up to 10 MB/s 3800 kS/s with NI-DAQmx, 1 MS/s with additional download. Special conditions apply.
Table 1. Channel, Speed, and Resolution Specifications
Output
Range (V)
±10
±10
±10
±10
Digital
I/O
82
82
8
8
Counter/
Timers
2, 24-bit
2, 24-bit
2, 24-bit
2, 24-bit
Triggers
Analog, digital
Analog, digital
Analog, digital
Analog, digital
Overview and Applications
National Instruments S Series devices combine the latest in PC
technologies to deliver simultaneous-sampling, high-bandwidth
capabilities for low and high-channel-count data acquisition and control
systems. Use S Series devices in a variety of applications including:
⢠Stimulus/response applications where synchronization among high-
speed analog input (AI), analog output (AO), and digital I/O is required
⢠42 V automotive applications
⢠Radar, sonar, ultrasound
⢠Transient/ballistics measurements
⢠High-energy physics
⢠IF digitization
Features
The NI high-performance S Series DAQ devices have the same
basic features of a typical multifunction device â eight digital I/O lines,
two counter/timers, two analog outputs, RTSI or PXI trigger bus for
multidevice synchronization, and PFI lines for software-controlled routing
of timing and triggering signals. S Series devices exceed the analog input
performance of traditional multiplexed devices with features including:
Onboard Memory
Family
(samples)
NI 6120
64 M
NI 6115 32 M or 64 M
NI 6110
8,192
NI 6111
8,192
1Filter can be bypassed.
Antialias Filters1
5-pole Bessel,
100 kHz
3-pole Bessel
50 and 500 kHz
â
â
Table 2. Special Features
Signal
Coupling
AC and DC
AC and DC
DC
DC
Input
Configuration
DIO
Pseudodifferential Hardware-timed
Pseudodifferential Hardware-timed
Pseudodifferential Software-timed
Pseudodifferential Software-timed
|
▷ |