English
Language : 

MVAC250-48AFD Datasheet, PDF (1/9 Pages) Murata Power Solutions Inc. – 250W 3 x 5 High Density AC-DC Power Supply Converter
www.murata-ps.com
FEATURES
IEC60601 Ed.3 medical (2 x MOPP Pri-Sec)
EN60950 ITE safety approved
250W compact high density
3" x 5" standard footprint
High efficiency up to 94%
Remote sense
Remote On/Off, Power OK (MVAC250-xxAFx)
Universal AC input with active PFC
Less than 1U high – 1.4"
Convection cooled operation up to 170W
Isolated 12V@1A fan output
Isolated 5V@2A standby/auxiliary output with
models MVAC250-xxAFx
RoHS compliant
Active inrush protection
Current sharing option
DESCRIPTION
The MVAC250 series switching power supplies
utilize advanced component and circuit
technologies to deliver high efficiency. Designed
for medical, computing, communications, telecom
and other OEM applications to satisfy 1U height
design considerations, the MVAC250 Series
measures only 3.0" x 5.0" x 1.40". All models
offer universal AC input with active power factor
correction (PFC) and compliance to worldwide
safety and EMC standards.
www.murata-ps.com/en/3d/acdc.html
www.murata-ps.com/en/3d/acdc.html
www.murata-ps.com/en/3d/acdc.html
www.murata-ps.com/en/3d/acdc.html
www.murata-ps.com/en/3d/acdc.html
www.murata-ps.com/en/3d/acdc.html
Available now at
www.murata-ps.com/en/3d/acdc.html
Ḥ
For full details go to
www.murata-ps.com/rohs
www.murata-ps.com/support
MVAC250 Series
250W 3" x 5" High Density AC-DC Power Supply Converter
ORDERING GUIDE
Model Number
MVAC250-12F
Natural Convection Cooling
Forced Air Cooling
Main Output Fan Output Aux Output
(V1)
(V2)
(V3)
12V
12V
MVAC250-24F
24V
12V
MVAC250-48F
50V
12V
MVAC250-12AF
MVAC250-12AFD*
170W
250W @ 250LFM
12V
12V
12V
12V
5V
5V
MVAC250-24AFD*
24V
12V
5V
MVAC250-48AFD*
50V
12V
5V
MVAC250-24AFT#
24V
12V
5V
MVAC-COVER
Optional cover kit assembly; see MVAC-COVER datasheet for details
* Refer to page 2 for current sharing model number MVAC250-xxAFD notes.
# CCC Certification is not available for these models.
INPUT CHARACTERISTICS
Parameter
Conditions
Min. Typ. Max. Units
Input Voltage Operating Range
Single phase
DC
90 115/230 264
Vac
127
300
Vdc
Input Frequency
47 50/60 63
Hz
Turn-on Input Voltage
Turn-off Input Voltage
Input rising
Input falling
80
70
90
80
Vac
Input Current
90Vac input, full load all outputs
3.4
A
No Load Input Power (MVAC250-xxAFD)7 (PS_ON = OFF, 5V_Aux = 0A)
1.5
2.0
W
Inrush Current
At 264Vac, at 25°C cold start
15
Apk
Power Factor
At 230Vac, full load
0.96
OUTPUT CHARACTERISTICS
Model Number
Main Output
Voltage (V1)
MVAC250-12F
12V
MVAC250-24F
MVAC250-24AFT
24V
MVAC250-48F
50V
MVAC250-12AF
12V
MVAC250-12AFD 12V @ 10.4A6
MVAC250-24AFD 24V @ 5.2A6
MVAC250-48AFD 50V @ 2.5A6
Load Current
0.4 to 20.8A
0.2 to 10.4A
0.1 to 5.0A
0 to 20.8A
0 to 20.8A
0 to 10.4A
0 to 5.0A
Maximum Load
Capacitance
0 to 1500μF
0 to 300μF
0 to 82μF
0 to 1500uF
0 to 1500μF
0 to 300μF
0 to 68μF
Line, Load, Cross
Regulation
± 1%
± 1%
± 1%
± 1%
± 1.5%6
± 1.5%6
+3.0% / –1.5%6
Typical Efficiency
@230Vac
93%
93%
94%
93%
93%
93%
94%
Main Output Characteristics (all models)
Parameter
Conditions
Transient Response9
50% load step, 1A/μsec slew rate
Settling Time to 1% of Nominal
Turn On Delay
After application of input power
Output Voltage Rise
Monotonic5
Output Holdup
120Vac/60Hz, full load
Temperature Coefficient
Ripple Voltage & Noise1
Compensates for up to 0.5V of lead drop with
Remote Sense
remote sense connected. Protected against
short circuit and reverse connection.
Typ.
Max. Units
±5
%
500
μsec
3
sec
20
50
msec
0.02 %/°C
1
%
500
mV
Auxiliary Output Characteristics (varies by model)
Auxiliary Output
Aux Output
Voltage8
Load Current
Load Capacitance
Fan (V2) all models
12V
0 to 1A
0 to 220μF
Aux (V3) – MVAC250-xxAFx
5V
0 to 2A
0 to 220μF
Line, Load, Cross Ripple Voltage &
Regulation3
Noise1
± 10%
2%
± 5%
1%
CB
Test Certificate
and Test Report
MVAC250.B02 Page 1 of 9