English
Language : 

MUU151 Datasheet, PDF (1/2 Pages) MPS Industries, Inc. – 150W U-bracket Switching Power Supplies For Medical Equipment
MUU151 SERIES
150W U-bracket Switching Power Supplies For Medical Equipment.
Description:
The MUU151 series of compact, open frame constructed, AC/DC
switching mode power supplies provide 150 Watts of continuous
output power.They are suited for use in hospital instrument and
many other applications. All models meet FCC Part-18 class B and
CISPR-11 EN55011 class B emission Limits and are designed to
comply with UL/c-UL(UL 60601-1:2ndEdition) and new CE
requirements. All units are 100% burned in and tested.
Features:
Wide Operating Voltage 100 to 240 VAC,47 to 63 Hz
Internal EMI filter
Single Output
Input connector mates with Molex housing 09-50-3031
and Molex 2478 series crimp terminal
Output connector mates with Molex housing 09-50-3081
and Molex 2478 series crimp terminal
Input Surge Current, Over Voltage and Over Load protection
Output Voltage Protection(Crowbar Design)
Active Power Factor Correction
Size: 3.21"x5"x1.66”
Class I
3 year warranty
Electrical Characteristics:
Safety Approvals :
CB
EN 60601-1
IEC 60601-1
Sym.
Vin
fin
PF
Po
Vo
Io
Iil
Iih
I rl
I rh
Eff
REG-i
REG -o
OVP
OCP
Ttr
Thold
Ts
Vp-p
Ilk
TC
Parameter
Test Conditions
Min. Typ. Max.
Input Voltage
Operating Voltage
100
240
Input Frequency
47
63
Power Factor Correction
Io=Full load, Vin=100~240VAC
0.95 0.97 1.0
Output Power Range
Vin=100 to 240 VAC
0
150
Output Voltage Range
See rating Chart
Output Current Range
See rating Chart
Input Current (Low Line) Io=Full load, Vin=115VAC
2
Input Current (High Line) Io=Full load, Vin=230VAC
1.2
Low Line Inrush Current
High Line Inrush Current
Efficiency
Io=Full load, 25 C,Cool start,
Vin=115VAC
Io=Full load, 25 C,Cool start,
Vin=230VAC
Io=Full load, Vin=230VAC
28 31
57 63
82 85 87
Line Regulation
Io=Full load
0.5 1
Load Regulation
Vin=230VAC
35
Over Voltage Protection
112
132
Over Current Protection
110
150
Time of Transient Response Io=Full load to Half Load, Vin=100VAC
4
Hold-Up Time
Io=Full load, Vin=110VAC
20
Start Up Time
Io=Full load, Vin=100VAC
2
Ripple & Noise (Peak to Peak) Full load, Vin=90VAC
0.5 1
Safety Ground Leakage Current Io=Full load, Vin=240VAC
0.075 0.1
Temperature Coefficient All output
-0.04
0.04
Unit
VAC
Hz
W
V
A
A
A
A
A
%
%
%
%
%
mS
mS
S
%
mA
%/ C
Environmental :
Sym.
Parameter
T e s t Conditions
Toper Operating Temperature
Tstg Storage Temperature
Ho Operating Humidity
Hr Storage Humidity
MTBF Operating Temperature at 25 C, Calculated per MIL-HDBK-217F
Pd Derate linearly from 100% load at 50 C to 50% load at 70 C
Min. Typ. Max. Unit
0 50 70
C
-40
85
C
0
95 %
0
75 %
0.1M
Hrs