English
Language : 

MBU120 Datasheet, PDF (1/2 Pages) MPS Industries, Inc. – 120W Opne Frame Switching Power Supplies For Medical Equipment.
MBU120 SERIES
120W Open Frame Switching Power Supplies For Medical Equipment.
Description:
The MBU120 series of compact, open frame constructed, AC/DC
switching mode power supplies provide 120 Watts of continuous output
power.They are suited for use in hospital instrument and many other
applications. All models meet FCC Part-18 class B and CISPR-11
EN55011 class B emission Limits and are designed to comply with
UL/c-UL(UL 60601-1:2ndEdition) ,TUV/T-mark(EN 60601-1:2ndEdition)
and new CE requirements. All units are 100% burned in and tested.
Features:
Wide Operating Voltage 90 to 260 VAC,47 to 63 Hz
Internal EMI filter
Single Output
Input connector mates with Molex housing 09-50-3051
and Molex 2478 series crimp terminal
Output Connector mates with screw terminal (Terminal Block)(16-22AWG)
or Molex housing 09-50-3121 and Molex 2478 series crimp terminal
Input Surge Current, Over Voltage and Over Load protection
Output Voltage Protection(Crowbar Design)
Active Power Factor Correction
Size: 3"x5"x1.42”
Class I
3 year warranty
Safety Approvals :
EN 60601-1
IEC 60601-1
Electrical Characteristics:
Sym.
Parameter
Test Conditions
Min. Typ. Max. Unit
Safety Approvals Input
Vin Voltage Range
100
240 VAC
Operate Voltage Range
90
260 VAC
fin Input Frequency
47
63 Hz
PF Power Factor Correction
Po Output Power Range
Io=Full load, Vin=90~260VAC
Vin=90 to 260 VAC
0.95 0.97 1.0
0
120 W
Vo Output Voltage Range
See rating Chart V
Io Output Current Range
See rating Chart A
Iil Input Current (Low Line) Io=Full load, Vin=115VAC
Iih Input Current (High Line) Io=Full load, Vin=230VAC
1.7 A
0.8 A
I rl Low Line Inrush Current
Io=Full load, 25 C,Cool start,
Vin=115VAC
48 54
A
I rh High Line Inrush Current
Io=Full load, 25 C,Cool start,
Vin=230VAC
58 63 A
Eff Efficiency
Io=Full load, Vin=230VAC
70 80 88 %
REG-i Line Regulation
Io=Full load
0.5 1
%
REG -o Load Regulation
Vin=230VAC
35
%
OVP Over Voltage Protection
112
132 %
OCP Over Current Protection
110
150 %
Ttr Time of Transient Response Io=Full load to Half Load, Vin=100VAC
4
mS
Thold Hold-Up Time
Io=Full load, Vin=110VAC
16
mS
Ts
Vp-p
Ilk
TC
Start Up Time
Io=Full load, Vin=100VAC
Ripple & Noise (Peak to Peak) Full load, Vin=90VAC
Safety Ground Leakage Current Io=Full load, Vin=240VAC
Temperature Coefficient All output
0.3 1 2
0.5 1
0.095 0.1
-0.04
0.04
S
%
mA
%/ C
Environmental :
Sym.
Parameter
T e s t Conditions
Toper Operating Temperature
Tstg Storage Temperature
Ho Operating Humidity
Hr Storage Humidity
MTBF Operating Temperature at 25 C, Calculated per MIL-HDBK-217F
Pd Derate linearly from 100% load at 50 C to 50% load at 70 C
Min. Typ. Max. Unit
0 50 70
C
-40
85
C
0
95 %
0
75 %
0.1M
Hrs