English
Language : 

HPU63B Datasheet, PDF (1/2 Pages) MPS Industries, Inc. – 63W Desk Top Swithching Power Supplies For health Care Applications
HPU63B SERIES
63W Desk Top Switching Power Supplies For Health Care Applications.
Description:
The HPU63B series of AC/DC switching mode power supplies
provide 63 Watts of continuous output power. All supplies are
UL94V-1 min compliant. They are suited for use in Health care
applications. All models meet FCC Part-18 class B and CISPR-11
EN55011 class B emission Limits and are designed to comply
with UL/c-UL(UL 60601-1:2ndEdition) ,TUV/T-mark (EN 60601-
1:2ndEdition), and new CE requirements. All units are 100%
burned in and tested.
Features:
Wide Operating Voltage 90 to 264 VAC,47 to 63 Hz
IEC-320-C8 Input Inlet
Single Output
Optional Output Connector (See appendix)
Input Surge Current, Over Voltage And Over Load protection
Output Voltage Protection(Crowbar Desing)
Energy Star 2.0, Efficiency level V
Class II
2 year warranty
Electrical Characteristics:
Safety Approvals :
EN 60601-1
IEC 60601-1
RT
PS
E
Agree to apply for the PSE
if order on hand
Sym.
Parameter
Test Conditions
Min. Typ. Max. Unit
Safety Approvals Input
Vin Voltage Range
100
240 VAC
fin
Po
Vo
Io
Iil
Iih
I rl
I rh
* Eff
REG-i
REG -o
OVP
OCP
Ttr
Thold
Ts
Vp-p
Ilk
Pno
TC
Operate Voltage Range
90
264 VAC
Input Frequency
47
63 Hz
Output Power Range
Vin=90 to 264VAC
0
63
W
Output Voltage Range
See rating chart V
Output Current Range
See rating chart A
Input Current (Low Line)
Input Current ( High Line)
Io=Full load, Vin=115 VAC
Io=Full load, Vin=230 VAC
1.62 A
0.72 A
Low Line Inrush Current
Io=Full load, 25 C ,Cool start,
Vin= 115VAC
26 30
A
High Line Inrush Current
Io=Full load, 25 C ,Cool start,
Vin= 230VAC
43 47 A
Efficiency
Io=Full Load, Vin=230VAC
87
92 %
Line Regulation
Io=Full Load
1
%
Load Regulation
Vin=230VAC
5
%
Over Voltage Protection
112
132 %
Over Current Protection
110
150 %
Time of Transient Response Io=Full Load to Half Load,Vin=100VAC
4
mS
Hold-Up Time
Io=Full Load, Vin=110VAC
12
mS
Start Up Time
Io=Full Load, Vi n= 10 0V AC
0.3 1 2
S
Ripple & Noise (Peak to Peak) Full Load, Vin=100VAC
1
%
Safety Ground Leakage C urrent Io = Fu ll Load, Vin=240VAC
No-Load Power Consumption No load,Vin=240VAC
0.1 mA
0.3 0.4 0.5 W
Temperature Coefficient
All output
-0.04
0.04 %/ C
* Note: The specifics for testing the energy efficiency of HPU63 Series are outlined in a separate document titled "Test Method for Calculating the
Energy Efficiency of Single-Voltage External Ac-Dc and Ac-Ac PowerSupplies (August 11, 2004)," which is available on the ENERGY STAR Website.
Environmental :
Sym.
Parameter
T e s t Conditions
Toper Operating Temperature
Tstg Storage Temperature
Ho Operating Humidity
Hr Storage Humidity
MTBF Operating Temperature at 25 C, Calculated per MIL-HDBK-217F
Pd Derate linearly from 100% load at 40 C to 50% load at 70 C
Min. Typ. Max. Unit
0 40 70
C
-40
85
C
0
95 %
0
75 %
0.1M
Hrs