English
Language : 

HPU31A Datasheet, PDF (1/2 Pages) MPS Industries, Inc. – HPU31A SERIES
HPU31A SERIES
30W Desk top type medical power supplies for Home health care applications
Description:
The HPU31A series of AC/DC switching mode power supplies
provide 30 Watts of continuous output power . All supplies are
UL94V-1 min compliant, include IEC-320-C18 input for worldwide
applications. They are suited for use in Home health care
applications. All models meet FCC Part-18 class B and CISPR-11
EN55011 class B emission Limits and are designed to comply with
UL/c-UL(UL 60601-1) ,TUV/T-mark (EN 60601-1) and new CE
requirements. All units are 100% burned in and tested.
Features:
Wide Input Voltage 90 to 264 VAC,47 to 63 Hz
IEC-320-C18 Input Inlet
Single Output
Output Voltage Available From 5VDC Thru 48VDC
Optional Output Connector (See appendix)
Over Voltage Protection (Crowbar Design)
Input Surge Current, Over Voltage And Over Load protection
CEC level V and Engergy Star 2.0 Compliance
Approved as Limited Power Source (LPS).
Class II
2 year warranty
Safety Approvals :
PS
E
Electrical Characteristics:
Sym.
Parameter
Test Conditions
Min. Typ. Max. Unit
Vin Input Voltage
Operating Voltage
90
264 VAC
fin Input Frequency
47
63 Hz
Po Output Power Range
Vin=90 to 264VAC
0
30
W
Vo Output Voltage Range
See rating chart V
Io Output Current Range
See rating chart A
Iil Input Current (Low Line)
Iih Input Current ( High Line)
Io=Full load, Vin=115 VAC
Io=Full load, Vin=230 VAC
0.9 A
0.34 A
I rl Low Line Inrush Current
Io=Full load, 25 C ,Cool start,
Vin= 115VAC
16 24
A
I rh High Line Inrush Current
Io=Full load, 25 C ,Cool start,
Vin= 230VAC
32 48 A
Eff Efficiency
Io=Full Load, Vin=230VAC
78 85 88 %
REG-i Line Regulation
Io=Full Load
0.5
%
REG -o Load Regulation
Vin=230VAC
5
%
OVP Over Voltage Protection
OCP Over Current Protection
112
132 %
110
150 %
Ttr Time of Transient Response Io=Full Load to Half Load,Vin=100VAC
4
mS
Thold Hold-Up Time
Ts Start Up Time
Io=Full Load, Vin=110VAC
Io=Full Load, Vi n= 10 0V AC
12
mS
0.3 1.5 2
S
Vrn Ripple & Noise (Peak to Peak) Full Load, Vin=90VAC
1
%
TC Temperature Coefficient
All output
-0.04
0.04 %/ C
Environmental :
Sym.
Parameter
Test Conditions
Toper Operating Temperature
Tstg Storage Temperature
Hr Relative Humidity
MTBF Operating Temperature at 25 C, Calculated per MIL-HDBK-217F
Pd Derate linearly from 100% load at 40 C to 50% load at 70 C
Min. Typ. Max.
0 40 70
-40
85
5
95
0.1M
Unit
C
C
%
Hrs
Safety Specifications:
Sym.
Parameter
Vps
Dielectric Withstanding Voltage
for Primary to secondary
CISPR EMI requirements for CISPR-11
FCC EMI requirements for FCC PART-18
Test Conditions
Primary to secondary
Vin=220VAC
Vin=110VAC
Min. Typ. Max. Unit
5600
VDC
B
CLASS
B
CLASS