English
Language : 

MPM-24M Datasheet, PDF (1/4 Pages) MicroPower Direct, LLC – 24W Output Power
MPM-24M Series
Medical Approved, 24W
Single & Dual Output
AC/DC Power Supplies
Key Features:
• 24W Output Power
• EN 60601-1 3RD Edition
• Universal 85-264 VAC Input
• EN 60950 Approved
• Reinforced Insulation
• Meets IEC Safety Class II
• Low Leakage Current
• Single and Dual Outputs
• Meets UL 508
Chassis Mount Models
& DIN Rail Mount
Option Available!
MicroPower Direct
292 Page Street
Suite D
Stoughton, MA 02072
USA
T: (781) 344-8226
F: (781) 344-8481
E: sales@micropowerdirect.com
W: www.micropowerdirect.com
Electrical Specifications
Specifications typical @ +25°C, 230 VAC input voltage & rated output current, unless otherwise noted. Specifications subject to change without notice.
Input
Parameter
Conditions
Min. Typ. Max. Units
Input Voltage Range
85
264
VAC
120
370
VDC
Input Frequency
47
440
Hz
Inrush Current, See Note 2
115 VAC
20
230 VAC
40
A Pk
No Load Power Consumption
0.3
W
Output
Parameter
Conditions
Min. Typ. Max. Units
Output Voltage Accuracy
±2.0
%
Line Regulation
VIN = Min to Max
±0.5
%
Load Regulation, IO = Min to Max
Single Output
±0.5
Dual Output
±2.5
%
Ripple & Noise (20 MHz), See Note 3
5.0 VDC Output
All Other Outputs
1.5
1.0
1.8
1.3
%VP-P of VO
Hold-Up Time
115 VAC, 60 Hz
20
230 VAC, 50 Hz
80
mS
Temperature Coefficient
±0.02
%/°C
Over Voltage Protection
See Note 4
120
% of VO
Overshoot
5
%
Overload Protection
See Note 5 105
% INOM
Over Temperature Protection
See Note 6
90
°C
Short Circuit Protection, See Note 7
Continuous (Autorecovery)
General
Parameter
Conditions
Min. Typ. Max. Units
Isolation Voltage (Reinforced)
Input to Output 4,000
VAC rms
Leakage Current
80
µA
Isolation Resistance
500 VDC 1,000
M⍀
Switching Frequency
132
kHz
Environmental
Parameter
Standard
Criterea
Level
Radiated Emissions
EN 55014, EN 55024
Class B
Conducted Emissions
EN 55022
Class B
ESD
EN 61000-4-2
A
A
±8 kV Air
±4 kV Contact
RS
EN 61000-4-3
A
10 V/m
EFT
EN 61000-4-4
A
±2 kV
Surge
EN 61000-4-5
A
±1 kV
CS
EN 61000-4-6
A
10 Vrms
PFM
EN 61000-4-8
A
30 A/M
Voltage Dips
EN 61000-4-11
A
30% 10 mS
Interruptions
EN 61000-4-11
B
95% 5,000 mS
Environmental
Parameter
Conditions
Min. Typ. Max. Units
Operating Temperature Range
Ambient -40 +25 +80
°C
Storage Temperature Range
-40
+95
°C
Cooling, See Note 8
Free Air Convection (See Derating Curve)
Humidity
RH, Non-condensing
95
%
Physical
Case Size
See Mechanical Diagrams (Pages 3 & 4)
Case Material
Non-Conductive Plastic Resin (UL94-V0)
Reliability Specifications
Parameter
Conditions
Min. Typ. Max. Units
MTBF
MIL HDBK 217F, 25°C, Gnd Benign 400
kHours
Safety Standards
UL 60950, EN 60950
Safety Class
IEC 60536 Class II
www.micropowerdirect.com