English
Language : 

GSM120B Datasheet, PDF (2/3 Pages) Mean Well Enterprises Co., Ltd. – 120WAC-DC High Reliability Medical Adaptor
120W AC-DC High Reliability Medical Adaptor
GSM120B series
SPECIFICATION
ORDER NO.
GSM120B12-R7B
GSM120B15-R7B
GSM120B20-R7B
GSM120B24-R7B
SAFETY MODEL NO.
GSM120B12
GSM120B15
GSM120B20
GSM120B24
DC VOLTAGE
Note.2 12V
15V
20V
24V
RATED CURRENT
8.5A
7A
6A
5A
CURRENT RANGE
0 ~ 8.5A
0 ~ 7A
0 ~ 6A
0 ~ 5A
RATED POWER (max.)
102W
105W
120W
120W
OUTPUT
RIPPLE & NOISE (max.) Note.3 120mVp-p
VOLTAGE TOLERANCE Note.4 ±5.0%
LINE REGULATION Note.5 ±1.0%
LOAD REGULATION
±5.0%
150mVp-p
±5.0%
±1.0%
±5.0%
180mVp-p
±5.0%
±1.0%
±4.0%
200mVp-p
±3.0%
±1.0%
±3.0%
SETUP, RISE TIME Note.6 1500ms, 30ms / 230VAC 2000ms, 30ms / 115VAC at full load
HOLD UP TIME (Typ.)
20ms / 230VAC 20ms / 115VAC at full load
VOLTAGE RANGE Note.7 80 ~ 264VAC 113 ~ 370VDC
FREQUENCY RANGE
47 ~ 63Hz
POWER FACTOR (Typ.)
PF>0.93 / 230VAC PF>0.97 / 115VAC at full load
INPUT EFFICIENCY (Typ.)
88%
89%
89%
90%
AC CURRENT (Typ.)
INRUSH CURRENT (Typ.)
LEAKAGE CURRENT(max.)
1.4A / 115VAC 0.7A / 230VAC
35A / 115VAC 70A / 230VAC
Touch current < 100 A/264VAC
OVERLOAD
105 ~ 160% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
PROTECTION
OVER VOLTAGE
105 ~ 135% rated output voltage
Protection type : Shut down o/p voltage, re-power on to recover
OVER TEMPERATURE
WORKING TEMP.
WORKING HUMIDITY
ENVIRONMENT STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
SAFETY STANDARDS
SAFETY & ISOLATION LEVEL
EMC
WITHSTAND VOLTAGE
(Note. 8) ISOLATION RESISTANCE
Shut down o/p voltage, re-power on to recover
-30 ~ +60℃ (Refer to "Derating Curve")
20% ~ 90% RH non-condensing
-40 ~ +85℃, 10 ~ 95% RH
±0.03% / ℃ (0 ~ 40℃)
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
ANSI/AAMI ES60601-1 / ES60601-1-11, TUV EN60601-1 / EN60601-1-11 approved
Primary-Secondary: 2xMOPP
I/P-O/P: 4KVAC
I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
EMC EMISSION
Compliance to EN55011(CISPR11) class B, EN61000-3-2,3, FCC PART 15 class B
EMC IMMUNITY
Compliance to EN61000-4-2,3,4,5,6,8,11, EN55024, EN60601-1-2, EN61204-3 medical level, criteria A
OTHERS
CONNECTOR
MTBF
DIMENSION
PACKING
PLUG
CABLE
372K hrs min. MIL-HDBK-217F(25℃)
167*67*35mm (L*W*H)
0.6Kg; 20pcs/13,0Kg/0.9CUFT
See page 3 ; Other type available by customer requested
See page 3 ; Other type available by customer requested
GSM120B48-R7B
GSM120B48
48V
2.5A
0 ~ 2.5A
120W
240mVp-p
±2.5%
±1.0%
±2.5%
91.5%
NOTE
1. All parameters are specified at 230VAC input, rated load, 25℃ 70% RH ambient.
2. DC voltage: The output voltage set at point measure by plug terminal & 50% load.
3. Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1uf & 47uf capacitor.
4. Tolerance: includes set up tolerance, line regulation, load regulation.
5. Line regulation is measured from low line to high line at rated load.
6. Length of set up time is measured at first cold start. Turning ON/OFF the power supply may lead to increase of the set up time.
7. Derating may be needed under low input voltage. Please check the derating curve for more details.
8. The power supply is considered as an independent unit, but the final equipment still need to re-confirm that the whole system complies with the
EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”
(as available on http://www.meanwell.com)
File Name:GSM120B-SPEC 2015-11-02