RPT-160 Datasheet, PDF (1/2 Pages) Mean Well Enterprises Co., Ltd. – 160W Triple Output Medical Type
160W Triple Output Medical Type
RPT-160 series
Features :
Universal AC input / Full range
Built-in active PFC function
Protections: Short circuit / Overload / Over voltage / Over temperature
5"x3" compact size
Free air convection for 100W and 145W with 20.5 CFM forced air
With power good and fail signal output
No load power consumption under 0.75W by PS-ON control (G model)
Standby 5V@0.8A with fan, @0.6A without fan (G model)
3 years warranty
G: With 5Vsb & no load power
RPT G
consumption < 0.75 W
Blank: Basic function (without 5Vsb)
160A
SPECIFICATION
EN60601-1
IEC60601-1
MODEL
RPT -160A
RPT -160B
RPT -160C
RPT -160D
OUTPUT NUMBER
DC VOLTAGE
CH1 CH2 CH3 CH1 CH2 CH3 CH1 CH2 CH3 CH1 CH2 CH3
5V
12V
-5V
5V
12V
-12V 5V
15V
-15V 5V
12V
24V
RATED CURRENT (20.5CFM) 14A
5.5A 1A
14A
5A
1A
14A
3.6A 1A
11A
5A
1.2A
CURRENT RANGE (convection) 0.6 ~ 9A 0.2 ~ 3.8A 0.1 ~ 0.6A 0.6 ~ 9A 0.2 ~ 3.4A 0.1 ~ 0.8A 0.6 ~ 9A 0.1 ~ 2.6A 0.1 ~ 0.8A 0.3 ~ 8A 0.2 ~ 2.6A 0.15 ~ 1A
CURRENT RANGE (20.5CFM) 0.6 ~ 14A 0.2 ~ 5.5A 0.1 ~ 1A 0.6 ~ 14A 0.2 ~ 5A 0.1 ~ 1A 0.6 ~ 14A 0.1 ~ 3.6A 0.1 ~ 1A 0.3 ~ 11A 0.2 ~ 5A 0.15 ~ 1.2A
OUTPUT
RATED POWER (convection) Note.7 98.6W
98.4W
99W
98.2W
RATED POWER (20.5CFM) Note.8 145W
146W
143W
147.8W
RIPPLE & NOISE (max.) Note.2 100mVp-p 120mVp-p 120mVp-p 100mVp-p 120mVp-p 120mVp-p 100mVp-p 150mVp-p 150mVp-p 100mVp-p 120mVp-p 200mVp-p
VOLTAGE ADJ. RANGE
CH1:5 ~ 5.5V
VOLTAGE TOLERANCE Note.3 2.0% 5.0% -5,+7% 2.0% 5.0% -4,+5% 2.0% 4.0% +8.0% 2.0% 5.0% +7,-5%
LINE REGULATION
0.5% 1.0% 1.0% 0.5% 1.0% 1.0% 0.5% 1.0% 1.0% 0.5% 1.0% 1.0%
LOAD REGULATION
1.5% 3.0% -5,+6% 1.5% 3.0% -4,+5% 2.0% 3.0% 8.0% 1.5% 3.0% -3,+4%
SETUP, RISE TIME
1800ms, 30ms/230VAC 3500ms, 30ms/115VAC at full load
HOLD UP TIME (Typ.)
16ms/230VAC/115VAC at full load
VOLTAGE RANGE Note.6 90 ~ 264VAC 127 ~ 370VDC
FREQUENCY RANGE
47 ~ 63Hz
POWER FACTOR (Typ.)
PF>0.93/230VAC PF>0.98/115VAC at full load
INPUT EFFICIENCY (Typ.)
84%
84%
83%
83%
AC CURRENT (Typ.)
INRUSH CURRENT (Typ.)
1.8A/115VAC 0.9A/230VAC
COLD START 35A/115VAC 70A/230VAC
LEAKAGE CURRENT
Earth leakage current <300uA / 264VAC, Patient leakage current <100uA/264VAC
OVERLOAD
105 ~ 135% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
OVER VOLTAGE
PROTECTION
OVER TEMPERATURE
CH1: 5.75 ~ 6.75V
Protection type : Shut down o/p voltage, re-power on to recover
105 (TSW1)detect on heatsink of power transistor
90 (TSW2 )detect on heatsink of power transistor
Protection type : (TSW1)Shut down o/p voltage, recovers automatically after temperature goes down
FUNCTION
Protection type : (TSW2)Shut down o/p voltage, re-power on to recover
5V STANDBY (G model)
5VSB : 5V@0.6A without fan, 0.8A with fan 20.5CFM ; tolerance 2%, ripple : 50mVp-p(max.)
PS-ON INPUT SIGNAL (G model) Power on: PS-ON = "Hi" or " > 2 ~ 5V" ; Power off: PS-ON = "Low" or " < 0 ~ 0.5V"
POWER GOOD / POWER FAIL 500ms>PG>10ms PF>1ms
WORKING TEMP.
-20 ~ +70 (Refer to "Derating Curve")
WORKING HUMIDITY
20 ~ 90% RH non-condensing
ENVIRONMENT STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
SAFETY STANDARDS
SAFETY & WITHSTAND VOLTAGE
EMC
ISOLATION RESISTANCE
(Note 4) EMC EMISSION
-40 ~ +85 , 10 ~ 95% RH
0.03%/ (0 ~ 50 )
10 ~ 500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes
UL60601-1, TUV EN60601-1 approved
I/P-O/P:4KVAC I/P-FG:1.5KVAC O/P-FG:0.5KVAC
I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25 / 70% RH
Compliance to EN55011 (CISPR11), EN55022 (CISPR22) Class B, EN61000-3-2,-3
OTHERS
NOTE
EMC IMMUNITY
Compliance to EN61000-4-2,3,4,5,6,8,11, EN55024, EN60601-1-2, EN61204-3, medical level, criteria A
MTBF
191.4Khrs min. MIL-HDBK-217F (25 )
DIMENSION
127*76.2*34.6mm (L*W*H)
PACKING
0.33Kg; 36pcs/12.9Kg/0.79CUFT
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets
EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”
(as available on http://www.meanwell.com)
5. HS1,HS2 & HS3 can not be shorted.
6. Derating may be needed under low input voltages. Please check the derating curve for more details.
7. The rated power includes 5Vsb @ 0.6A.
8. The rated power includes 5Vsb @ 0.8A.