RPS-160_12 Datasheet, PDF (1/2 Pages) Mean Well Enterprises Co., Ltd. – 160W Single Output Medical Type
160W Single Output Medical Type
RPS-160 series
Features :
Universal AC input / Full range
Built-in active PFC function
Protections: Short circuit / Overload / Over voltage / Over temperature
5"x3" compact size
Free air convection for 110W and 160W with 20.5 CFM forced air
With power good and fail signal output
Built-in remote sense function for 5~15V
No load power consumption under 0.75W by PS-ON control (G model)
Standby 5V@0.8A with fan, @0.6A without fan (G model)
3 years warranty
G: With 5Vsb & no load power
RPS G
consumption < 0.75 W
Blank: Basic function (without 5Vsb)
160 12
SPECIFICATION
MODEL
RPS -160-5
RPS -160-12
RPS -160-15
RPS -160-24
RPS -160-48
DC VOLTAGE
5V
12V
15V
24V
48V
RATED CURRENT (20.5CFM) 30A
12.9A
10.3A
6.5A
3.25A
CURRENT RANGE (convection) 0 ~ 20A
0 ~ 9.1A
0 ~ 7.3A
0 ~ 4.6A
0 ~ 2.3A
CURRENT RANGE (20.5CFM) 0 ~ 30A
0 ~ 12.9A
0 ~ 10.3A
0 ~ 6.5A
0 ~ 3.25A
RATED POWER (convection) Note.7 103W
112.2W
112.5W
113.4W
113.4W
OUTPUT
RATED POWER (20.5CFM) Note.8 155W
RIPPLE & NOISE (max.) Note.2 100mVp-p
VOLTAGE ADJ. RANGE
CH1:4.5 ~ 5.5V
VOLTAGE TOLERANCE Note.3 4.0%
LINE REGULATION
0.5%
159.8W
100mVp-p
CH1:10.8 ~ 13.2V
3.0%
0.5%
159.5W
120mVp-p
CH1:13.5 ~ 16.5V
3.0%
0.5%
161W
150mVp-p
CH1:22 ~ 27V
2.0%
0.5%
161W
250mVp-p
CH1:43.2 ~ 52.8V
2.0%
0.5%
LOAD REGULATION
1.0%
1.0%
1.0%
1.0%
1.0%
SETUP, RISE TIME
1800ms, 30ms/230VAC 3500ms, 30ms/115VAC at full load
HOLD UP TIME (Typ.)
16ms/230VAC/115VAC at full load
VOLTAGE RANGE Note.6 90 ~ 264VAC 127 ~ 370VDC
FREQUENCY RANGE
47 ~ 63Hz
POWER FACTOR (Typ.)
PF>0.93/230VAC PF>0.98/115VAC at full load
INPUT EFFICIENCY (Typ.)
85%
87%
87%
87%
88%
AC CURRENT (Typ.)
INRUSH CURRENT (Typ.)
2A/115VAC 1.1A/230VAC
COLD START 35A/115VAC 70A/230VAC
LEAKAGE CURRENT
Earth leakage current <300uA / 264VAC, patient leakage current <100uA/264VAC
OVERLOAD
105 ~ 135% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
OVER VOLTAGE
PROTECTION
OVER TEMPERATURE
5.75 ~ 6.75V
13.8 ~ 16.2V
17.25 ~ 20.25V
27.6 ~ 32.4V
Protection type : Shut down o/p voltage, re-power on to recover
110 (5V),105 (12V,15V,24V,48V) (TSW1 : detect on heatsink of power transistor)
95 (5V),90 (12V,15V,24V,48V) (TSW2 : detect on heatsink of power transistor)
Protection type : (TSW1)Shut down o/p voltage, recovers automatically after temperature goes down
55.2 ~ 64.8V
Protection type : (TSW2)Shut down o/p voltage, re-power on to recover
5V STANDBY (G model)
5VSB : 5V@0.6A without fan, 0.8A with fan 20.5CFM ; tolerance 2%, ripple : 50mVp-p(max.)
PS-ON INPUT SIGNAL (G model) Power on: PS-ON = "Hi" or " > 2 ~ 5V" ; Power off: PS-ON = "Low" or " < 0 ~ 0.5V"
FUNCTION POWER GOOD / POWER FAIL 500ms>PG>10ms
PF>1ms
REMOTE SENSE
5 ~ 15V
WORKING TEMP.
-20 ~ +70 (Refer to "Derating Curve")
WORKING HUMIDITY
20 ~ 90% RH non-condensing
ENVIRONMENT STORAGE TEMP., HUMIDITY -40 ~ +85 , 10 ~ 95% RH
TEMP. COEFFICIENT
0.03%/ (0 ~ 50 )
VIBRATION
10 ~ 500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes
SAFETY STANDARDS
ANSI/AAMI ES60601-1, TUV EN60601-1 approved
SAFETY & WITHSTAND VOLTAGE
EMC
ISOLATION RESISTANCE
(Note 4) EMC EMISSION
I/P-O/P:4KVAC I/P-FG:1.5KVAC O/P-FG:0.5KVAC
I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25 / 70% RH
Compliance to EN55011 (CISPR11), EN55022 (CISPR22) Class B, EN61000-3-2,-3
OTHERS
NOTE
EMC IMMUNITY
Compliance to EN61000-4-2,3,4,5,6,8,11, EN55024, EN60601-1-2, EN61204-3, medical level, criteria A
MTBF
230.5Khrs min. MIL-HDBK-217F (25 )
DIMENSION
127*76.2*34.6mm (L*W*H)
PACKING
0.32Kg; 36pcs/12.5Kg/0.79CUFT
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets
EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”
(as available on http://www.meanwell.com)
5. HS1,HS2 & HS3 can not be shorted.
6. Derating may be needed under low input voltages. Please check the derating curve for more details.
7. The rated power includes 5Vsb @ 0.6A.
8. The rated power includes 5Vsb @ 0.8A.