RPD-75_12 Datasheet, PDF (1/2 Pages) Mean Well Enterprises Co., Ltd. – 75W Dual Output Medical Type
75W Dual Output Medical Type
RPD-75 series
Features :
Universal AC input/Full range
Low leakage current<200uA
Protections: Short circuit / Overload / Over voltage
Free air convection for rated power and 23.5CFM forced air
convection for peak load
ANSI/AAMI ES60601-1/IEC60601-1/EN60601-1 medical safety approved
Fixed switching frequency at 65KHz
3 years warranty
SPECIFICATION
MODEL
RPD-75A
RPD-75B
OUTPUT NUMBER
CH1
CH2
CH1
CH2
DC VOLTAGE
5V
12V
5V
24V
RATED CURRENT
7A
3A
5A
2A
CURRENT RANGE
1 ~ 9.5A
0.3 ~ 4A
1 ~ 6.8A
0.2 ~ 2.7A
OUTPUT
RATED POWER
71W
PEAK LOAD (23.5CFM)
95.5W
RIPPLE & NOISE (max.) Note.2 80mVp-p
VOLTAGE ADJ. RANGE
CH1: 4.75 ~ 5.5V
VOLTAGE TOLERANCE Note.3 2.0%
LINE REGULATION
0.5%
120mVp-p
6.0%
1.0%
73W
98.8W
80mVp-p
CH1: 4.75 ~ 5.5V
2.0%
0.5%
200mVp-p
6.0%
1.0%
LOAD REGULATION
1.5%
3.0%
1.5%
3.0%
SETUP, RISE TIME
500ms, 30ms/230VAC 500ms, 30ms/115VAC at full load
HOLD UP TIME (Typ.)
80ms/230VAC 20ms/115VAC at full load
VOLTAGE RANGE
90 ~ 264VAC 127 ~ 370VDC
FREQUENCY RANGE
47 ~ 63Hz
INPUT
EFFICIENCY (Typ.)
77%
79%
AC CURRENT (Typ.)
1.5A/115VAC 1A/230VAC
INRUSH CURRENT (Typ.) COLD START 25A/115VAC 50A/230VAC
LEAKAGE CURRENT
Earth leakage current <200uA / 264VAC, Patient leakage current <100uA/264VAC
OVERLOAD
PROTECTION
OVER VOLTAGE
140 ~ 180% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
CH1: 5.75 ~ 6.75V
Protection type : Shut down o/p voltage, re-power on to recover
WORKING TEMP.
WORKING HUMIDITY
ENVIRONMENT STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
SAFETY STANDARDS
SAFETY & WITHSTAND VOLTAGE
EMC
ISOLATION RESISTANCE
(Note 4) EMC EMISSION
-20 ~ +70 (Refer to "Derating Curve")
20 ~ 90% RH non-condensing
-40 ~ +85 , 10 ~ 95% RH
0.03%/ (0 ~ 45 )
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
ANSI/AAMI ES60601-1, TUV EN60601-1, IEC60601-1 approved
I/P-O/P:4KVAC I/P-FG:1.5KVAC O/P-FG:1.5KVAC
I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25 / 70% RH
Compliance to EN55011 (CISPR11), EN55022 (CISPR22) Class B, EN61000-3-2,-3
EMC IMMUNITY
Compliance to EN61000-4-2,3,4,5,6,8,11, EN55024, EN60601-1-2, EN61000-6-2, EN61204-3, heavy industry level, EN61204-3 medical level, criteria A
OTHERS
NOTE
MTBF
569.9K hrs min. MIL-HDBK-217F (25 )
DIMENSION
127*76.2*31mm (L*W*H)
PACKING
0.25Kg; 63pcs/16.7Kg/1.35CUFT
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets
EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”
(as available on http://www.meanwell.com)
5. Length of set up time is measured at cold first start. Turning ON/OFF the power supply may lead to increase of the set up time.
6. Heat Sink HS1,HS2,HS3 can not be shorted.
File Name:RPD-75-SPEC 2012-02-20