RPD-60 Datasheet, PDF (1/2 Pages) Mean Well Enterprises Co., Ltd. – 60W Dual Output Medical Type
60W Dual Output Medical Type
RPD-60 series
Features :
4"x2" miniature size
Universal AC input/Full range
Low leakage current<200uA
Protections: Short circuit / Over load / Over voltage
Cooling by free air convection
UL60601-1/IEC60601-1/EN60601-1 medical safety approved
UL60950-1/IEC60950-1/EN60950-1 ITE safety approved
Fixed switch frequency at 100KHz
3 years warranty
SPECIFICATION
EN60601-1
IEC60601-1
MODEL
RPD-60A
RPD-60B
OUTPUT
INPUT
OUTPUT NUMBER
CH1
CH2
DC VOLTAGE
5V
12V
RATED CURRENT
5A
2A
CURRENT RANGE
0.5 ~ 5.5A
0.1 ~ 2.2A
RATED POWER
49W
PEAK LOAD(10sec.) Note.4 53.9W
RIPPLE & NOISE (max.) Note.2 80mVp-p
80mVp-p
VOLTAGE TOLERANCE Note.3 +3,-2%
6.0%
LINE REGULATION
0.5%
1.0%
LOAD REGULATION
1.5%
2.0%
SETUP, RISE TIME
200ms, 15ms/230VAC 200ms, 15ms/115VAC at full load
HOLD UP TIME (Typ.)
70ms/230VAC 14ms/115VAC at full load
VOLTAGE RANGE
90 ~ 264VAC 127 ~ 370VDC
FREQUENCY RANGE
47 ~ 63Hz
EFFICIENCY (Typ.)
78%
AC CURRENT (Typ.)
1.1A/115VAC 0.7 A/230VAC
INRUSH CURRENT (Typ.) COLD START 60A/230VAC 30A/115VAC
CH1
5V
3.5A
0.5 ~ 3.85A
53.5W
58.85W
80mVp-p
+3,-2%
0.5%
1.5%
82%
CH2
24V
1.5A
0.1 ~ 1.65A
100mVp-p
+8,-4%
1.0%
2.0%
LEAKAGE CURRENT
For earth <200uA/264VAC, For patient <100uA/264VAC
OVERLOAD
PROTECTION
OVER VOLTAGE
115 ~ 150% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
CH1: 5.75 ~ 6.75V
Protection type : Shut down o/p voltage, re-power on to recover
WORKING TEMP.
-20 ~ +65 (Refer to output load derating curve)
WORKING HUMIDITY
20 ~ 90% RH non-condensing
ENVIRONMENT STORAGE TEMP., HUMIDITY -40 ~ +85 , 10 ~ 95% RH
TEMP. COEFFICIENT
0.03%/ (0 ~ 45 )
VIBRATION
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
SAFETY STANDARDS
UL60950-1, TUV EN60950-1, UL60601-1, TUV EN60601-1, IEC60601-1 approved
SAFETY &
EMC
(Note 5)
WITHSTAND VOLTAGE
ISOLATION RESISTANCE
EMI CONDUCTION & RADIATION
HARMONIC CURRENT
I/P-O/P:4KVAC I/P-FG:1.5KVAC O/P-FG:1.5KVAC
I/P-O/P, I/P-FG, O/P-FG:100M Ohms/500VDC
Compliance to EN55011(CISPR11),EN55022 (CISPR22) Class B
Compliance to EN61000-3-2,-3
EMS IMMUNITY
Compliance to EN61000-4-2,3,4,5,6,8,11; ENV50204, EN55024, EN60601-1-2, EN61204-3, medical level, criteria A
MTBF
677.8Khrs min. MIL-HDBK-217F (25 )
OTHERS DIMENSION
101.6*50.8*29mm (L*W*H)
PACKING
0.15Kg; 96pcs/15.4Kg/0.89CUFT
NOTE
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. 33% Duty cycle maximum within every 30 seconds. Average output power should not exceed the rated power.
5. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets
EMC directives.
6. Length of set up time is measured at cold first start. Turning ON/OFF the power supply may lead to increase of the set up time.
File Name:RPD-60-SPEC 2008-03-25