English
Language : 

MSP-200-24 Datasheet, PDF (1/4 Pages) Mean Well Enterprises Co., Ltd. – 200W Single Output Medical Type
200W Single Output Medical Type
MSP-200 series
■ Features :
‧Universal AC input / Full range
‧Built-in active PFC function, PF>0.95
‧High efficiency up to 89%
‧Withstand 300VAC surge input for 5 seconds
‧Protections: Short circuit / Overload / Over voltage / Over temperature
‧Cooling by free air convection
‧Built-in constant current limiting circuit
‧1U low profile 38mm
‧Medical safety approved (MOOP level)
‧Built-in remote ON-OFF control
‧Standby 5V@0.3A
‧Built-in remote sense function
‧No load power consumption<0.5W (Note.6)
‧5 years warranty
SPECIFICATION
MODEL
MSP-200-3.3 MSP-200-5 MSP-200-7.5 MSP-200-12 MSP-200-15 MSP-200-24 MSP-200-36 MSP-200-48
DC VOLTAGE
3.3V
5V
7.5V
12V
15V
24V
36V
48V
RATED CURRENT
CURRENT RANGE
RATED POWER
40A
0 ~ 40A
132W
35A
0 ~ 35A
175W
26.7A
0 ~ 26.7A
200.3W
16.7A
0 ~ 16.7A
200.4W
13.4A
0 ~ 13.4A
201W
8.4A
0 ~ 8.4A
201.6W
5.7A
0 ~ 5.7A
205.2W
4.3A
0 ~ 4.3A
206.4W
RIPPLE & NOISE (max.) Note.2 80mVp-p
90mVp-p
100mVp-p 120mVp-p 150mVp-p 150mVp-p 250mVp-p 250mVp-p
OUTPUT
VOLTAGE ADJ. RANGE
2.8 ~ 3.8V
VOLTAGE TOLERANCE Note.3 ±2.0%
LINE REGULATION
±0.5%
LOAD REGULATION
±1.5%
4.3 ~ 5.8V
±2.0%
±0.5%
±1.0%
6.8 ~ 9V
±2.0%
±0.5%
±1.0%
10.2 ~ 13.8V
±1.0%
±0.3%
±0.5%
13.5 ~ 18V
±1.0%
±0.3%
±0.5%
21.6 ~ 28.8V
±1.0%
±0.2%
±0.5%
28.8 ~ 39.6V
±1.0%
±0.2%
±0.5%
40.8 ~ 55.2V
±1.0%
±0.2%
±0.5%
SETUP, RISE TIME
HOLD UP TIME (Typ.)
1000ms, 50ms/230VAC 2500ms, 50ms/115VAC at full load
16ms/230VAC 16ms/115VAC at full load
VOLTAGE RANGE Note.5 85 ~ 264VAC
FREQUENCY RANGE
47 ~ 63Hz
120 ~ 370VDC
POWER FACTOR (Typ.)
PF>0.95/230VAC PF>0.99/115VAC at full load
INPUT EFFICIENCY (Typ.)
80%
84%
86%
88%
88%
88%
89%
89%
AC CURRENT (Typ.)
INRUSH CURRENT (Typ.)
2.2A/115VAC 1.1A/230VAC
35A/115VAC 70A/230VAC
LEAKAGE CURRENT Note.7 Earth leakage current < 300μA/264VAC , Touch leakage current < 100μA/264VAC
OVERLOAD
PROTECTION
OVER VOLTAGE
105 ~ 135% rated output power
Protection type : Constant current limiting, recovers automatically after fault condition is removed
3.96 ~ 4.62V 6 ~ 7V
9.4 ~ 10.9V 14.4 ~ 16.8V 18.8 ~ 21.8V 30 ~ 34.8V 41.4 ~ 48.6V
Protection type : Shut down o/p voltage, re-power on to recover
57.6 ~ 67.2V
OVER TEMPERATURE
Shut down o/p voltage, recovers automatically after temperature goes down
5V STANDBY
FUNCTION
REMOTE CONTROL
5VSB : 5V@0.3A ; tolerance±5%, ripple : 50mVp-p(max.)
RC+/RC- : 4 ~10V or open = power on ; 0 ~ 0.8V or short = power off
WORKING TEMP.
-40 ~ +70℃ (Refer to "Derating Curve")
WORKING HUMIDITY
20 ~ 90% RH non-condensing
ENVIRONMENT STORAGE TEMP., HUMIDITY -40 ~ +85℃, 10 ~ 95% RH
TEMP. COEFFICIENT
±0.03%/℃ (0 ~ 50℃)
VIBRATION
10 ~ 500Hz, 5G 10min./1cycle, 60min. each along X, Y, Z axes
SAFETY STANDARDS
ANSI/AAMI ES60601-1, IEC60601-1 approved
SAFETY & WITHSTAND VOLTAGE
EMC
ISOLATION RESISTANCE
(Note 4) EMC EMISSION
I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:0.5KVAC
I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25℃/ 70% RH
Compliance to EN55011 (CISPR11) Class B, EN61000-3-2,-3
EMC IMMUNITY
MTBF
Compliance to EN61000-4-2,3,4,5,6,8,11, EN60601-1-2
209.4K hrs min. MIL-HDBK-217F (25℃)
OTHERS DIMENSION
PACKING
199*98*38mm (L*W*H)
0.77Kg; 18pcs/14.9Kg/0.9CUFT
NOTE
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25℃ of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets
EMC directives. For guidance on how to perform these EMC tests, please refer to EMI testing of component power supplies.
(as available on http://www.meanwell.com)
5. Derating may be needed under low input voltages. Please check the derating curve for more details.
6. No load power consumption<0.5W when RC+ & RC- (CN100 pin1,2) 0 ~ 8V or short.
7. Touch current was measured from primary input to DC output.
File Name:MSP-200-SPEC 2014-02-12