English
Language : 

6EHT1 Datasheet, PDF (1/2 Pages) Tyco Electronics – High Performance RFI Power Line Filters for Medical Equipment
HT Series
CORCOM Product Guide
High Performance RFI Power Line Filters for Medical Equipment
HT6
UL Recognized
CSA Certified
VDE Approved
HT7
HT3
HT1
HT Series
The HT series power line filters provide superior common-
mode and premium differential-mode attenuation to RFI
noise on the frequency range from 10kHz to 30MHz, for
IEC 60601 and UL2601 patient-connected equipment, as
well as all other electronic equipment that cannot tolerate
leakage current.
They are a size- and cost-effective solution to the problem
of meeting both medical leakage current limits and
conducted emissions limits in equipment with high noise
sources (such as switching power supplies). HT series
filters can meet the following Conducted Emissions
standards:
• European Normalized Standards based on EN55022,
Level A (in many cases, Level B also).
• FCC Part 15, Computing Devices, Class B.
• FCC Part 18, ISM Equipment, Industrial and
Consumer.
Note: When using the models with an IEC connector,
remember that the leakage current of the companion line
cord, GA400, is more significant than that of the filter -
which may disqualify this line cord from use in patient care
applications.
For more information on UL2601, see Appendix C.
Electrical Schematics
3 & 6 Amp
LINE
LOAD
10 Amp
LINE
LOAD
Specifications
Maximum leakage current,each line-to-ground
@ 120 VAC 60 Hz:
@ 250 VAC 50 Hz:
Hipot rating (one minute):
line-to-ground
line-to-line
Operating frequency:
2µA
5µA
1500 VAC
1450 VDC
50/60 Hz
Rated voltage (max.):
Minimum insertion loss in dB:
250 VAC
Line-to-ground in 50 ohm circuit
Current
Frequency–MHz
Rating .01 .02 .05 .08 .15 .5 1 2 5 10 20 30
3A 22 32 36 - 49 46 40 30 22 12 12 12
6A 16 23 32 41 46 41 33 26 15 9 6 2
10A 9 15 24 30 36 42 34 22 11 12 8 8
15A 4 9 18 22 27 41 34 22 12 12 5 2
Line-to-line in 50 ohm circuit
Current
Frequency–MHz
Rating .01 .02 .05 .08 .15 .5 1 2 5 10 20 30
3A 3 1 30 - 61 70 65 65 48 40 32 32
6A 4 1 14 45 51 70 70 65 55 47 37 37
10A 7 8 17 32 52 70 70 70 65 55 40 35
15A 12 16 15 10 51 70 70 70 70 70 65 55
15 Amp
LINE
LOAD
Resistor location for reference only.
70
Catalog 1654001
Revised 05-07
Dimensions are in inches and
millimeters unless otherwise
specified. Values in italics
are metric equivalents.
Dimensions are shown for
reference purposes only.
Specifications subject
to change.
USA Cust. Svc.: 1-800-468-2023
CORCOM Prods: 1-847-680-7400
Germany:
49-89-6089-0
www.corcom.com
www.tycoelectronics.com