9 | Keysight | U1270 Series Handheld Digital Multimeters - Data Sheet
Specification Assumptions
–– Accuracy is given as ±(% of reading + counts of least significant digit) at 23 °C ± 5 °C, with relative humidity less than
80% RH.
–– AC V and AC μA/mA/A specifications are AC coupled, true RMS and are valid from 5% of range to 100% of range.
–– The crest factor may be up to 3.0 at full scale except for the 1000 V range where it is 1.5 at full scale.
–– For non-sinusoidal waveforms, add (2% reading + 2% full scale) typical, for crest factors up to 3.
–– After ZLOW voltage measurements, wait at least 20 minutes for thermal impact to cool before proceeding with any
other measurement.