English
Language : 

N1913A Datasheet, PDF (9/28 Pages) Keysight Technologies – EPM Series Power Meters E-Series and 8480 Series Power Sensors
09 | Keysight | N1913A and N1914A EPM Series Power Meters / E-Series and 8480 Series Power Sensors - Data Sheet
N1913A/14A EPM Series Power Meters Performance Characteristics (continued)
Power meter functions
Accessed by key entry
Either hard keys, or soft key menu, and programmable
Zero
Zeros the meter. (Power reference calibrator is switched off during zeroing.)
Cal
Calibrates the meter using internal (power reference calibrator) or external source. Reference cal factor settable from 1%
to 150%, in 0.1% increments.
Frequency
Entered frequency range is used to interpolate the calibration factors table. Frequency range from 1 kHz to 999.9 GHz.
Also settable in 1 kHz steps.
Cal factor
Sets the calibration factor for the meter. Range: 1% to 150%, in 0.1% increments.
Relative
Displays all successive measurements relative to the last displayed value
Offset
Allows power measurements to be offset by –100 dB to +100 dB, settable in 0.001 dB increments, to compensate for
external loss or gain
Save/recall
Store up to 10 instrument states via the save/recall menu
dBm/W
Selectable units of either Watts or dBm in absolute power; or percent or dB for relative measurements
Filter (averaging)
Selectable from 1 to 1024. Auto-averaging provides automatic noise compensation.
Duty cycle
Duty cycle values between 0.001% to 99.999%, in 0.001% increments, can be entered to display a peak power repre-
sentation of measured power. The following equation is used to calculate the displayed peak power value: peak power =
measured power/duty cycle.
Sensor cal tables
Selects cal factor versus frequency tables corresponding to specified sensors
Limits
High and low limits can be set in the range –150.000 dBm to +230.000 dBm, in 0.001 dBm increments
Preset default values
dBm mode, rel off, power reference off, duty cycle off, offset off, frequency 50 MHz, AUTO average, free run, AUTO range
(for E-Series sensors and N8480 Series)
Display
Color display with selectable single and split screen formats are available. A quasi-analog display is available for peaking
measurements. The dual channel power meter can simultaneously display any two configurations of A, B, A/B, B/A, A-B, B-A
and relative. With the optional USB ports, additional dual channel (C & D), adds up to total 4-channels measurement display.
Power meter general specifications
Dimensions
The following dimensions exclude front and rear protrusions:
212.6 mm W x 88.5 mm H x 348.3 mm D (8.5 in x 3.5 in x 13.7 in)
Weight
Model
N1913A
N1914A
Net
Shipping
3.6 kg (8.0 lb) 8.2 kg (18.1 lb)
3.7 kg (8.2 lb) 8.2 kg (18.3 lb)
Rear panel connectors
Recorder outputs
Analog 0 to 1 volt, 1 kΩ output impedance, BNC connector. N1914A recorder outputs are dedicated to channel A and
channel B.
GPIB, USB 2.0 and
10/100BaseT LAN
Trigger Input (optional)1
Trigger Output (optional)1
Interfaces to allow communication with an external controller
Input has TTL compatible logic levels and uses a BNC connector.
High: > 2.4 V
Low: < 0.7 V
Output provides TTL compatible logic levels and uses a BNC connector.
High: > 2.4 V
Low: < 0.7 V
Ground
Binding post, accepts 4 mm plug or bare wire connection
USB Host (options)
USB ports which connects to U2000 series USB power sensors
VGA Out (options)
Standard 15-pin VGA connector, allows connection of external VGA monitor
1.  For automated power or frequency sweep function