English
Language : 

E6705A Datasheet, PDF (7/8 Pages) Keysight Technologies – N8993A GS-8800 Series (8960) Wireless
07 | Keysight | N8993A GS-8800 Series (8960) Wireless Communication Design Verification (DV) Bench-top System - Data Sheet
Specifications
Generic instrument specification
E5515E general specification
GPIB
Dimensions (H x W x D)
Weight
Display
LAN (local area network)
LAN port
(for firmware upgrades only)
LAN 2 port
(for high data throughput)
LAN 3 port (for future use)
Calibration interval
IEEE Standard 488.2
235 mm x 425 mm x 629 mm (9.25 in x 16.7 in x 24.7 in),
7 rack spaces high
31.9 kg (70.1 lbs)
26.7 cm (10.5 in), active matrix, color, liquid crystal
RJ-45 connector, 100 Mbps
RJ-45 connector, 1000 Mbps (for E5515E)
RJ-45 connector, 1000 Mbps
Two years
EMI
EMI conducted and radiated interference meets CISPR-11
PC/peripheral input/output
USB
Serial (RS-232)
LAN
GPIB
Six auxiliary ports are available: two on the front panel and four on
the rear panel of the industrial PC
Two on the on the rear panel of the industrial PC, DB9 male connector
Three 100 Base-T Ethernet, RJ-45 connector
One on the rear panel of the industrial PC
Timebase specification
Internal timebase
External timebase
Internal timebase of the 8960/E5515E wireless communications test
set is fed to all other instruments in the system. The 8960’s internal
oven-controlled crystal oscillator’s specifications are as follows:
–– Aging rates: < ±0.1 ppm per year, < ±0.005 ppm peak-to-peak
per day during any 24-hour period starting 24 hours or more
after a cold start
–– Temperature stability: < ±0.01 ppm, frequency variation from
+25 °C over the temperature range 0 to +55 °C
Warm-up time: 5 minutes to be within ±0.1 ppm of frequency at one
hour, 15 minutes to be within ±0.01 ppm of frequency at one hour
When the external reference signal is present on the EXT REF IN, the
system instruments will be locked to it
General specifications
Operating conditions
General
Storage temperature
Operating temperature
Accuracy specified
temperature
Humidity (relative)
Altitude
Power requirement
Indoor
–20 to +70 °C
+10 to +30 °C
+20 to +30 °C (Refer to measurement accuracy specification of
individual radio technology for further information)
5 to 80% relative humidity (non-condensing)
0 to 2 km
90 to 254 VAC, 50 to 60 Hz, 4118 VA maximum