AC Specifications (continued)
AC Operating Characteristics:
Maximum reading rates 0.01% of ac step additional error. Additional settling delay required when input dc level varies.
Function
ACV and ACI
ACV and ACI
ACV and ACI
ACV and ACI
ACV and ACI
Digits
6.5
6.5
6.5
6.5
6.5
Reading/s
7 s/reading
1
1.61
10
502
AC Filter
Slow
Medium
Fast
Fast
Fast
1 For External Trigger or remote operation using default settling delay (Delay Auto).
2 Maximum useful limit with default settling delays used.
Systems Speeds:
Maximum useful limit with default settling delays used; Speeds are for 4.5 digits, Delay 0, and Fast ac filter.
Function or range change:
Autorange time:
ASCII reading to GPIB:
Max. internal trigger rate:
Max. external trigger rate to memory:
5/s
<0.8 s
50/s
50/s
50/s
Additional Low-Frequency Errors (% of reading):
Input >100 mV. For mV input, multiply % of reading error x 10.
Frequency
3 Hz-5 Hz
5 Hz-10 Hz
10 Hz-40 Hz
40 Hz-100 Hz
100 Hz-300 Hz
300 Hz-1 kHz
>1 kHz
6.5 digits
0
0
0
0
0
0
0
5.5 digits
0.12
0.17
0.2
0.06
0.03
0.01
0
4.5 digits
0.12
0.17
0.2
0.21
0.21
0.07
0.02
7