M9183A Datasheet, PDF (6/14 Pages) Keysight Technologies – PXI Digital Multimeters 6½ Digit, High Performance
06 | Keysight | PXI Digital Multimeters, 6½ Digit, High Performance (M9182A, M9183A) - Data Sheet
Technical Specifications and Characteristics (continued)
M9182A and M9183A: Accuracy specifications ±(% of reading + % of range)1,2
Function
Range
Full scale reading or
resolution
24 hour
23 °C ± 1 °C
Single shot
duty cycle11
[M9183A only]
2 - 100 Hz
100 Hz - 1 kHz
1 - 10 kHz
0.02 %
0.20 %
2.00 %
0.03 ± 0.03
0.03 ± 0.30
0.03 ± 3.00
Single shot
14 μs - 62.5 ms
1 µs
pulse width12, 14
[M9183A only]
.01 ± 4 µs
Capacitance13
[M9183A and
M9182A]
1000.0 pF
10.000 nF
100.00 nF
1.0000 µF
10.000 µF
100.00 µF
1.0000 mF
10.000 mF
1199.9 pF
11.999 nF
119.99 nF
1.1999 µF
11.999 µF
119.99 µF
1.1999 mF
11.999 mF
1.00 + 0.10
1.20 + 0.05
1.00 + 0.10
1.00 + 0.10
1.00 + 0.10
1.00 + 0.10
1.20 + 0.10
2.00 + 0.10
90 day
23 °C ± 5 °C
0.03 ± 0.03
0.03 ± 0.30
0.03 ± 3.00
01 ± 4 µs
1.00 + 0.10
1.20 + 0.05
1.00 + 0.10
1.00 + 0.10
1.00 + 0.10
1.00 + 0.10
1.20 + 0.10
2.00 + 0.10
1 year
23 °C ± 5 °C
0.03 ± 0.03
0.03 ± 0.30
0.03 ± 3.00
01 ± 4 µs
1.00 + 0.10
1.20 + 0.05
1.00 + 0.10
1.00 + 0.10
1.00 + 0.10
1.00 + 0.10
1.20 + 0.10
2.00 + 0.10
1.  Specifications are for 1 hour warm up, within 1 hour self-cal, aperture ≥ 0.5 sec. Slow AC filter for AC measurements only.
2.  For temperatures outside the range of 23 °C ± 5 °C, but within 0 °C to 50 °C, add 0.1 × accuracy specification per °C.
3.  20% over range on all ranges except 300 V range, 10% over range for 300 V range.
4.  Minimum input specified: 5 mV or 1% of range, whichever is larger.
5.  Signal is limited to 8x106 Volt Hz product. For example, at 32 kHz, the highest input is 250 V.
6.  For inputs from 5 mV to 15 mV, add 100 µV to the specification.
7.  Specifications are for 4-wire resistance measurements, for 2-wire, add 1 mΩ additional error to the specification; for offset compensated
ohms (M9183A only), add 0.02% of range.
8.  Minimum input specified: 60 µA or 1.5% of range, whichever is larger; for inputs < 5% of full scale, add 0.02% of range.
9.  For inputs from 60 to 120 µA, add 10 µA to the specification; for inputs < 5% of full scale, add 0.1% of range.
10.  Minimum amplitude greater of: 100 mV, or 5 % of range for 1 Hz to 2.5 kHz, or 25 % of range for 2.5 kHz to 200 kHz , or 40% of range for
200 kHz to 300 kHz..
11.  Specifications are % of reading (0.03) ± adder.
12.  Specifications are % of reading + time.
13.  Specifications apply to input signals ≥ 5% of range.
14.  Maximum wait time for duty cycle and period is 5 seconds.
Definitions for Specifications
Specification (spec): Represents warranted performance of a calibrated instrument that
has been stored for a minimum of two hours within the operating temperature range of
0 to 50 °C, unless otherwise stated, and after a one hour warm-up period. The specifications
include measurement uncertainty. Data represented in this document are specifications
unless otherwise noted.
Typical (typ): Represents characteristic performance, which 80% of the instruments manu-
factured will meet. This data is not warranted, does not include measurement uncertainty,
and is valid only at room temperature (approximately 25 °C).
Nominal (nom): The expected mean or average performance, or an attribute whose perfor-
mance is by design, such as the 50 Ω connector. This data is not warranted and is measured
at room temperature (approximately 25 °C).
Measured (meas): An attribute measured during the design phase for purposes of commu-
nicating expected performance, such as amplitude drift vs. time. This data is not warranted
and is measured at room temperature (approximately 25 °C).
Note: All graphs contain measured data from several units at room temperature unless
otherwise noted.