English
Language : 

N9030A Datasheet, PDF (5/28 Pages) Keysight Technologies – PXA X-Series Signal Analyzer
5 | Keysight | N9030A PXA X-Series Signal Analyzer - Data Sheet
Sweep time and triggering
Range
Accuracy
Sweep trigger
Trigger Delay
Time gating
Gate methods
Gate length range (except method = FFT)
Gate delay range
Gate delay jitter
Sweep (trace) point range
All spans
Resolution bandwidth (RBW)
Range (–3.01 dB bandwidth)
Bandwidth accuracy (power)
RBW range
Bandwidth accuracy (–3.01 dB)
RBW range
Selectivity (–60 dB/–3 dB)
EMI bandwidth (CISPR compliant)
EMI bandwidth (MIL STD 461E compliant)
Analysis bandwidth 1
Maximum bandwidth
Video bandwidth (VBW)
Range
Accuracy
Measurement speed 2
Local measurement and display update rate
Remote measurement and LAN transfer rate
Marker peak search
Center frequency tune and transfer (RF)
Center frequency tune and transfer (μW)
Measurement/mode switching
Span = 0 Hz
Span ≥ 10 Hz
1 μs to 6000 s
1 ms to 4000 s
Span ≥ 10 Hz, swept
Span ≥ 10 Hz, FFT
Span = 0 Hz
± 0.01% nominal
± 40% nominal
± 0.01% nominal
Free run, line, video, external 1, external 2, RF burst, periodic timer
Span = 0 Hz or FFT
Span ≥ 10 Hz, swept
Resolution
–150 to +500 ms
0 to 500 ms
0.1 μs
Gated LO; gated video; gated FFT
1 μs to 5.0 s
0 to 100.0 s
33.3 ns p-p nominal
1 to 40001
1 Hz to 3 MHz (10% steps), 4, 5, 6, 8 MHz
1 Hz to 100 kHz
110 kHz to 1.0 MHz (< 3.6 GHz CF)
1.1 to 2 MHz (< 3.6 GHz CF)
2.2 to 3 MHz (< 3.6 GHz CF)
4 to 8 MHz (< 3.6 GHz CF)
± 0.5% (± 0.022 dB)
± 1.0% (± 0.044 dB)
± 0.07 dB nominal
± 0.10 dB nominal
± 0.20 dB nominal
1 Hz to 1.3 MHz
200 Hz, 9 kHz, 120 kHz, 1 MHz
10 Hz, 100 Hz, 1 kHz, 10 kHz,
100 kHz, 1 MHz
± 2% nominal
4.1:1 nominal
(Option EMC required)
(Option EMC required)
Standard
Option B25
Option B40
Option B85
Option B1X
10 MHz
25 MHz
40 MHz
85 MHz
160 MHz
1 Hz to 3 MHz (10% steps), 4, 5, 6, 8 MHz, and wide open (labeled 50 MHz)
± 6% nominal (in swept mode and zero span)
Standard
10 ms (100/s) nominal
10 ms (100/s) nominal
2.5 ms nominal
43 ms nominal
69 ms nominal
40 ms nominal
1.  Analysis bandwidth is the instantaneous bandwidth available around a center frequency over which the input signal can be digitized for fur-
ther analysis or processing in the time, frequency, or modulation domain.
2.  Sweep points = 101.