04 | Keysight | U1240 Series Handheld Digital Multimeters - Data Sheet
Electrical Specifications
AC specifications
Function
Range
Resolution
Test current/burden voltage Accuracy ± (% of reading + counts of least significant digit)
AC voltage1, 5 1000.0 mV 0.1 mV
â
True RMS
10.000 V
0.001 V
â
100.00 V
0.01 V
â
40 to 500 Hz
1% + 5
500 Hz to 1 kHz
2% + 5
1% + 5
1 to 2 kHz
â
2% + 5
1000.0 V
0.1 V
â
â
AC current2, 5
True RMS
1000.0 µA
10000 µA
0.1 µA
1 µA
< 0.06 V (50 â¦)
< 0.55 V (50 â¦)
100.00 mA
440.0 mA3
10.000 A4
0.01 mA
0.1 mA
0.001 A
< 0.18 V (0.5 â¦)
< 0.8 V (0.5 â¦)
< 0.4 V (0.01 â¦)
1% + 5
1.5% + 5
â
1.â Input impedance: 10 MΩ (nominal) in parallel with < 100 pF, with overload protection of 1000 V RMS
2.â Crest factor ⤠3. For non-sinusoidal waveforms with crest factor up to 3, add 2% reading + 2% full scale typical.
3.â Current can be measured from 50 mA to 440 mA continuously. An additional 0.2% needs to be added to the specified accuracy if the signal
measured is in the range of 440 mA to 1100 mA for 30 seconds maximum. After measuring a current of > 440 mA, leave the meter to cool
down for twice the measuring time used before application of low current measurement.
4.â Current can be measured from 0.5 A up to 10 A continuously with a maximum operating temperature of 50 °C. An additional 0.3% needs
to be added to the specified accuracy if the signal measured is in the range of 10 A to 19.999 A for 15 seconds maximum. After measuring a
current of >10 A, leave the meter to cool down for 60 seconds before applying a low current measurement.
5.â AC voltage and AC current specifications are AC coupled. True RMS measurement is valid from 5 % of range to 100 % of range.
Temperature specifications
Thermocouple type
K (for U1241B and U1242B)
J (for U1242B only)
Range
â40 to 1000 ºC/â48 to 1832 ºF
â40 to 1000 ºC/â48 to 1832 ºF
Resolution
0.1 °C /0.1 °F
0.1 ºC/0.1 ºF
Accuracy ± (% of reading + offset error)
1% + 1 ºC/1% + 1.8 °F
1% + 1 ºC/1% + 1.8 °F
Capacitance specifications
Range
1000.0 nF
10.000 µF
100.00 µF
1000.0 µF
10.000 mF
Resolution
0.1 nF
0.001 µF
0.01 µF
0.1 µF
0.001 mF
Accuracy ± (% of reading + counts of least significant digit)
1.2% + 4
2% + 4
Harmonic ratio specifications
Range
0.0% to 99.9%
Frequency
40 to 500 Hz
Voltage
100 mVAC to 1000 VAC
|
▷ |