English
Language : 

N9912A Datasheet, PDF (31/34 Pages) Keysight Technologies – N9912A FieldFox RF Handheld Analyzer
General Information (continued)
External reference/trigger in
Connector
Input frequency
Input amplitude range
Lock range
Trigger input
Impedance
Level Range
BNC (f)
10 MHz
–5 dBm to +10 dBm, 50 Ω (nominal)
±10 ppm of external reference frequency (nominal)
10 kΩ (nominal)
Rising edge: 1.7 V (nominal)
Falling edge: 1 V (nominal)
FieldFox Data Link Software
FieldFox Data Link software, installed on a PC, provides the following capabilities:
- Capture of current trace and settings
- Opening of data files (s1p, s2p, csv, sta, and png) residing
on the instrument
- Editing cal kit and cable files on the instrument, or creating
new cal kits and cables
- Transferring files to/from the instrument
- Annotating plots for documentation purposes
- Marker, limit line, and format changes on the PC
- Report generation
- Printing function
FieldFox Data Link software is available from the following website:
http://www.keysight.com/find/fieldfoxsupport
31