English
Language : 

M9370A Datasheet, PDF (30/34 Pages) Keysight Technologies – Keysight PXIe Vector Network Analyzers
Multisite operation
Multi-site operation is the ability to configure multiple independent PXI VNAs to operate independently
in a single PXIe chassis. Up to ten independent PXI VNA instances have been demonstrated, allowing
parallel testing of devices. Each instance of an independent PXI VNA can have different number of
ports, and can be triggered synchronously, or asynchronously.
More than four independent PXI VNAs operating on a single controller may experience some
degradation in performance due to processing limitations.
Description
Data points
Measurements
Formats
Data display formats
Channels
Number of channels
IF bandwidths
Group delay
Aperture (selectable)
Maximum aperture
Minimum delay
Display range
Log magnitude S11 or S22
Log magnitude S21 or S12
Log magnitude resolution
Phase
Phase resolution
VSWR
Information
32,001 (using controller with 32-bit OS)
100,001 (using controller with 64-bit OS)
S11, S21, S12 and S22 magnitude and phase
Full S-parameter matrix for multiport operation
Log magnitude, linear magnitude, VSWR, phase, group delay, impedance, real and
imaginary
Up to 200 independent measurement channels. A measurement channel is
coupled to stimulus response settings including frequency, IF bandwidth, power
level and number of points.
10 Hz, 20 Hz, 30 Hz, 50 Hz, 100 Hz, 200 Hz, 300 Hz, 500 Hz, 1 kHz, 2 KHz, 3 kHz,
5 kHz, 10 kHz, 20 kHz, 30 kHz, 50 kHz, 100 kHz, 300 kHz, 600 kHz, 1.2 MHz
(frequency span)/(number of points –1)
25% of frequency span
Limited to measuring no more than 180° of phase change within the minimum
aperture.
-1000 to 1000 dB
-1000 to 1000 dB
0.01 dB
-180 to +180 degrees
0.01 degrees
1.01 to 1000
30