English
Language : 

E5053A Datasheet, PDF (3/22 Pages) Keysight Technologies – E5052B Signal Source Analyzer
03 | Keysight | E5052B Signal Source Analyzer 10 MHz to 7 GHz, 26.5 GHz, or 110 GHz - Data Sheet
Phase Noise Measurement
Table 1-2. Phase noise measurement performance
Description
Specification (E5052B)
Specification (E5052B Option 011)
RF IN frequency range
10 MHz to 7 GHz
Measurement frequency bands
10 MHZ to 41 MHz, 39 MHz to 101 MHz,
99 MHz to 1.5 GHz, 250 MHz to 7 GHz1
RF frequency tracking range
0.4% of carrier frequency
Measurement parameters
SSB phase noise [dBc/Hz], Spurious noise [dBc],
Integrated rms phase deviation [deg, rad] or time jitter [s],
Residual FM [Hz rms]
Number of trace
1 data trace and 1 memory trace with ‘data math’ functions
Measurement trigger
continuous/single/hold source: internal/external/manual/bus
Offset frequency range (effective)
RF carrier signal > 1 GHz
(> 400 MHz for wide capture mode)
1 Hz to 100 MHz
1 Hz to 40 MHz (wide capture mode)
10 Hz to 100 MHz
10 Hz to 40 MHz (wide capture mode)
RF carrier signal < 1 GHz
(< 400 MHz for wide capture mode)
1 Hz to 10% of carrier frequency
10 Hz to 10% of carrier frequency
Phase noise uncertainty2 at effective offset frequencies
Offset 1 Hz to 10 Hz
± 4 dB (SPD)
N/A
Offset 10 Hz to 100 Hz
± 4 dB (SPD)
Offset 100 Hz to 1 kHz
± 3 dB
Offset 1 kHz to 40 MHz
± 2 dB (± 3 dB for wide capture range mode)
Offset 40 MHz to 100 MHz
± 3 dB
SSB phase noise sensitivity
See Table 1-3, 1-4, 1-5, Figure 1-1, 1-2, 1-3
IF gain setting
0 dB to 50 dB in 10 dB step
( not available in wide capture mode)
0 dB to 30 dB in 10 dB step
(not available in wide capture mode)
Enhanced sensitivity
Cross-correlation method available.
N/A
Number of correlation = 1 to 10,000
See Table 1-5 and Figure 1-3
Built-in LO phase noise optimization
< 150 kHz (optimized for better close-in phase noise measurement)
> 150 kHz (optimized for better far-out phase noise measurement)
See Figure 1-4.
Reference oscillator bandwidth optimization
Narrow / Wide See Figure 1-5.
Residual spurious response level
< –80 dBc (SPD) at > 10 kHz offset frequency with correlation > 120 sec.
except for 23.5 MHz ± 1 MHz and 71 MHz ± 3 MHz of carrier frequency
< –65 dBc (typical) at 1 kHz to 10 kHz offset frequency
Measurement time
See Table 1-6
Measurement range
Capture mode: Normal or Wide
PN mode (Regular)
RBW: Auto
X-axis: Offset frequency in log scale
Segment PN mode3
RBW: 96 mHz to 25 kHz (stepped), Maximum N/A
offset
frequency span: 93.2 Hz to 24.4 MHz (stepped)
X-axis: Offset frequency in liner scale
1.  Wide capture mode is available for 250 MHz to 7 GHz only
2.  Phase noise uncertainty: specified at 10 MHz and 1 GHz of carrier frequency with 0 dBm level. PN level > –60 dBc
3.  Segment PN mode is available with the firmware revision 3.20 or later. In this datasheet, specification of phase noise sensitivity, residual spurious response
level and measurement time are applicable to the regular PN mode. In the segment PN mode, number of these parameters depend on the measurement
setting. For more detail, refer to the user’s manual.