English
Language : 

N9914A Datasheet, PDF (25/48 Pages) Keysight Technologies – FieldFox Handheld Analyzers
25 | Keysight | FieldFox Handheld Analyzers 4/6.5/9/14/18/26.5/32/44/50 GHz - Data Sheet
Spectrum Analyzer (continued)
Dynamic range specifications
Displayed average noise level (DANL) - (dBm)
Input terminated, RMS detection, log averaging, 0 dB input attenuation, reference level of -20 dBm, normalized to 1 Hz RBW
N991xA, N993xA1
Preamp off
Spec (23 ± 5 °C)
Spec (-10 to 55 °C)
Typical (23 ± 5 °C)
Typical (-10 to 55 °C)
2 MHz to 4.5 GHz2
-137
-135
-139
-138
> 4.5 to 7 GHz
-133
-131
-136
-130
> 7 to 13 GHz
-129
-127
-132
-130
> 13 to 17 GHz
-124
-122
-126
-125
> 17 to 22 GHz
-119
-117
-122
-121
> 22 to 25 GHz
-114
-111
-117
-114
> 25 to 26.5 GHz
-110
-108
-112
-111
Preamp on
Spec (23 ± 5 °C)
Spec (-10 to 55 °C)
Typical (23 ± 5 °C)
Typical (-10 to 55 °C)
2 MHz to 4.5 GHz2
-153
-151
-155
-154
> 4.5 to 7 GHz
-149
-147
-151
-150
> 7 to 13 GHz
-147
-145
-149
-148
> 13 to 17 GHz
-143
-141
-145
-144
> 17 to 22 GHz
-140
-139
-143
-142
> 22 to 25 GHz
-134
-132
-137
-134
> 25 to 26.5 GHz
-128
-126
-131
-129
N995xA, N996xA
Preamp off
Spec (23 ± 5 °C)
Spec (-10 to 55 °C)
Typical (23 ± 5 °C)
Typical (-10 to 55 °C)
9 kHz to 2 MHz
-91
-91
-118
-118
> 2 MHz to 2.1 GHz
-137
-135
-143
-141
> 2.1 to 2.8 GHz
-135
-133
-142
-140
> 2.8 to 4.5 GHz
-137
-135
-143
-141
> 4.5 to 7 GHz
-134
-133
-140
-138
> 7 to 13 GHz
-134
-132
-141
-139
> 13 to 22 GHz
-132
-129
-140
-137
> 22 to 35 GHz
-130
-127
-137
-134
> 35 to 40 GHz
-122
-119
-132
-129
> 40 to 46 GHz
-119
-116
-126
-123
> 46 to 50 GHz
-117
-112
-124
-120
Preamp on
Spec (23 ± 5 °C)
Spec (-10 to 55 °C)
Typical (23 ± 5 °C)
Typical (-10 to 55 °C)
9 kHz to 2 MHz
-94
-94
-131
-130
> 2 MHz to 2.1 GHz
-153
-151
-159
-158
> 2.1 to 2.8 GHz
-151
-149
-157
-155
> 2.8 to 4.5 GHz
-153
-151
-158
-156
> 4.5 to 7 GHz
-150
-149
-156
-154
> 7 to 13 GHz
-146
-144
-152
-150
> 13 to 22 GHz
-142
-139
-149
-147
> 22 to 35 GHz
-141
-139
-147
-145
> 35 to 40 GHz
-136
-132
-144
-141
> 40 to 46 GHz
-131
-128
-138
-135
> 46 to 50 GHz
-126
-123
-135
-132
1.  9 kHz to 2 MHz: -116 (nominal) preamp off, -120 (nominal) preamp on, applicable only for FieldFox with serial number prefixes of MY5607/SG5607/US5607
and FieldFox upgraded with Option N9910HU-100/200/300/400.
2.  2. Add 4 dB between 2.1 and 2.8 GHz.