English
Language : 

N5249A Datasheet, PDF (25/62 Pages) Keysight Technologies – Keysight 2-Port and 4-Port PNA-X Network Analyzer
Table 9f. Maximum Leveled Power (dBm), Option 224
Description
Specification
Typical
Source 2 Out 1
Source 2
Out 2
Source 2 Out 1
Source 2
Out 2
Filt. Mode2 Hi Pwr Mode2
Filt. Mode2 Hi Pwr Mode2
10 MHz to 50 MHz
9
18
13
12
21
18
50 MHz to 500 MHz
11
18
17
13
22
21
500 MHz to 3.2 GHz
10
14
14
13
17
19
3.2 GHz to 8.5 GHz
18
18
18
22
22
22
8.5 GHz to 10 GHz
18
18
18
22
22
22
10 GHz to 13.5 GHz
16
16
16
21
21
20
13.5 GHz to 16 GHz
16
16
16
21
21
20
16 GHz to 20 GHz
15
15
13
19
19
17
20 GHz to 24 GHz
13
13
12
18
18
17
24 GHz to 26.5 GHz
7
7
5
14
14
11
1 In Filtered Mode, the signal path goes through filters to minimize harmonics below 3.2 GHz. In Hi Power Mode, the signal
bypasses the filters to maximize output power.
Table 9g. Maximum Leveled Power (dB), Options 224 or 423 with 0291
Description
Specification
Typical
Port 1
Ports 2
Port 1
Ports 2
Filt. Mode2 Hi Pwr Mode2
Filt. Mode2
Hi Pwr Mode2
10 MHz to 50 MHz
6
12
13
8
18
17
50 MHz to 500 MHz
8
13
13
11
20
20
500 MHz to 3.2 GHz
8
10
12
11
13
16
3.2 GHz to 8.5 GHz
13
13
13
19
19
17
8.5 GHz to 10 GHz
13
13
13
19
19
17
10 GHz to 13.5 GHz
12
12
10
15
15
14
13.5 GHz to 16 GHz
12
12
10
15
15
14
16 GHz to 20 GHz
10
10
6
13
13
9
20 GHz to 24 GHz
6
6
4
11
11
8
24 GHz to 26.5 GHz
0
0
-2
8
8
4
1 Option 029 affects port 1 and port 2 maximum leveled power. Refer to Table 9c for other ports.
2 In Filtered Mode, the signal path goes through filters to minimize harmonics below 3.2 GHz. In Hi Power Mode, the signal
bypasses the filters to maximize output power.
25